A Dual-Modal Hybrid Marker for Automatic Registration and Coordinate Hand-Over in C-Arm CT-Guided Spinal Navigation
An asymmetric hybrid registration frame (RF) integrates radiopaque and optical markers to automate computed tomography (CT) registration and coordinate hand-over to a dynamic reference frame (DRF). We compared automatic and manual fiducial selection using three fixed CT datasets of a rigid sawbone phantom. Ten operator...