Sep 2026· Nuclear medicine communications· 0 citations· 9 references
Medicine
TL;DR
The SPECT/CT systems at NORI Hospital Islamabad met established NEMA standards for image quality and developed a structured quality control program based on acceptance testing results.
Abstract
Background
Single photon emission computed tomography/computed tomography (SPECT/CT) systems play a crucial role in nuclear medicine for accurate diagnosis and treatment planning. Ensuring high image quality through rigorous acceptance testing and quality control is essential to maintain diagnostic reliability and patient safety.
Objective
This study evaluated the performance of SPECT/CT systems at NORI Hospital Islamabad against National Electrical Manufacturers Association (NEMA) and International Atomic Energy Agency standards and developed a structured quality control program based on acceptance testing results.
Methods
Acceptance tests were carried out using Tc-99m, I-131, and Co-57 radioactive sources, in conjunction with quadrant bar and Jaszczak phantoms. Parameters assessed included spatial resolution, intrinsic and extrinsic uniformity, energy resolution, sensitivity, center of rotation (COR), detector shielding, background counts, leakage, CT number accuracy, and nuclear medicine/CT registration. Data were compared with manufacturer specifications and NEMA standards to evaluate system performance and inform quality control strategies.
Results
Both detectors exhibited intrinsic and extrinsic uniformity within NEMA limits, ensuring consistent image quality. Spatial resolution testing confirmed sharp and precise imaging, while energy resolution was within manufacturer-specified values. Sensitivity measurements showed minor differences between detectors but remained within acceptable ranges. COR values were precise (<±0.55 mm), confirming proper mechanical and software alignment. CT number accuracy and nuclear medicine/CT registration tests confirmed precise anatomical-functional image overlay. Detector shielding and leakage tests showed minimal interference, supporting reliable imaging.
Conclusion
The SPECT/CT systems at NORI Hospital Islamabad met established NEMA standards for image quality. Acceptance testing validated detector performance and highlighted minor areas for improvement, such as periodic sensitivity calibration. Implementation of a structured quality control program ensures consistent equipment performance, optimal image quality, improved diagnostic accuracy, and superior patient care. Routine adherence to international standards is essential for maintaining high-quality nuclear medicine imaging practices.
The Siemens Naeotom Alpha PCCT scanner satisfies the requirements noted in task group reports for a radiotherapy simulator and has the potential to offer advantages for radiation oncology applications, including improved target delineation and treatment planning accuracy.
P. Basran, D. Richtsmeier· Journal of Applied Clinical...· 0 citations
PURPOSE
The study evaluated the image quality of the cone beam computed tomography (CBCT) system on Varian TrueBeam linear accelerators with HyperSight (TB-HS) and compared it with the standard TrueBeam system (TB-S). Particular emphasis was placed on changes relevant to offline adaptive radiotherapy (ART).
METHODS
I...
J. Lohela, S. Rytky, S. Karhula et al.· Annals of Biomedical Enginee...· 0 citations
Background Computed tomography (CT) contributes substantially to medical radiation exposure, making diagnostic reference levels (DRLs) important for dose optimization. This study aimed to establish preliminary institutional DRLs and evaluate acceptable quality dose (AQD) at an Indian tertiary care center. Methods A cro...
Ashutosh Pandey, V. K. Dhingra, Pankaj Sharma· F1000Research· 0 citations
The findings support the use of optimized low dose CT protocols in SPECT/CT without compromising quantitative reliability, helping radiographers and nuclear medicine technologist balance radiation protection with diagnostic requirements.
E. Karić, M. Kukuljan, L. Lezaic et al.· Radiography· 0 citations
Core parameters of the Musculoskeletal Ultrasound Computed Tomography System (MUCTS), such as spatial resolution, slice thickness, and three-dimensional reconstruction volume error, directly determine imaging performance and clinical diagnostic accuracy. To objectively and systematically assess the imaging quality and...
Xiao-Xue Bai, Li Zhao, Pan Zhou et al.· Bioengineering· 0 citations
BACKGROUND
Head and neck computed tomography angiography (CTA) requires high spatial resolution while maintaining acceptable image noise. This study aimed to compare image quality between photon-counting CT (PCCT) and energy-integrating detector CT (EID-CT) and to evaluate PCCT reconstruction strategies for specific cl...
Wei Qiao, Xiao Lu, Wei Zhou et al.· European Journal of Radiolog...· 0 citations
Related blog posts
Microsoft Research Blog· microsoft.comOct 30, 2026
Evaluating Microsoft Copilot at enterprise scale requires test data that captures the full complexity of real organizations, yet real enterprise data contains sensitive information that cannot be used for evaluation under any circumstances. The post From synthetic enterprise data evaluations to real Microsoft Copilot AI confidence appeared first on Microsoft Research.
Microsoft Research Blog· microsoft.comOct 29, 2026
Host Sowbhagya Suresh shares some of the latest news and cutting-edge research from across Microsoft—from understanding the challenges facing coding agents to uncovering how people use AI at work and improving extended-range weather forecasting. The post Microsoft Research Focus – Edition #97 appeared first on Microsoft Research.
“X-Hinges” helps design and 3D print items like lamps and gloves with an interactive twist: automatically embedded sensors that perceive and adaptively respond to human contact.
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.