A competition-format microarray assay on a poly(oligo(ethylene glycol) methacrylate-co-glycidyl methacrylate-co-glycidyl methacrylate brush integrated into a single-use microfluidic cassette and portable reader to quantify vancomycin directly from fingerstick volumes of whole blood in 30 min.
Abstract
Summary Vancomycin requires individualized dosing guided by therapeutic drug monitoring, but current laboratory assays are slow and centralized. We describe a competition-format microarray assay on a poly(oligo(ethylene glycol) methacrylate-co-glycidyl methacrylate) brush integrated into a single-use microfluidic cassette and portable reader to quantify vancomycin directly from fingerstick volumes of whole blood in 30 min. The assay shows a limit of detection of 0.8 μg/mL and precise quantification across the clinical range, with relative error below 20%. In a pilot study of 31 patient samples, cassette measurements correlated strongly with routine hospital testing with minimal bias, including when operated by minimally trained coordinators. This work establishes an approach for rapid, decentralized vancomycin monitoring that could support area-under-the-curve-based dosing in settings where access to high-complexity laboratories is limited and suggests a path toward extending therapeutic drug monitoring to point-of-care devices.
This review aims to critically evaluate emerging bioanalytical platforms, modeling frameworks, and decision-support systems for optimizing TDM of last-resort antibiotics, and proposes a three-tiered TDM model combining site-specific sensing, real-time analysis, and computational forecasting to improve antimicrobial ste...
Aparna Inamdar, N. Beeraka, P. R. Hemanth Vikram et al.· Frontiers in Pharmacology· 1 citation
: Point-of-care molecular diagnostics (POC-MDx) have transformed infectious disease diagnosis by enabling rapid, accurate pathogen detection at or near the patient care site. Unlike conventional laboratory-based molecular tests, POC molecular platforms provide clinically actionable results within minutes to hours, faci...
Yogesh Kumar Gupta· International Journal of Sci...· 0 citations
A whole-blood multiplexed vertical flow assay (WB-xVFA) that performs single-tier Lyme serology directly from 50 microliter of whole blood is presented, demonstrating the feasibility of rapid, near-patient whole-blood Lyme serology.
Mihye Lee, R. Ghosh, Artem Goncharov et al.· 0 citations
A novel point-of-care test, the TAC-D4 POCT, capable of measuring tacrolimus levels directly from whole blood samples without sample preparation steps or expensive analyzers with automated results available in 1-hour is described, helping overcome the current logistical barriers in the clinical management of patients r...
Damon T. Burrow, David S. Kinnamon, J. Heggestad et al.· Analytical Chemistry· 1 citation
A battery-free, smartphone-connected electrochemical lateral flow assay with linked analytics for the detection of procalcitonin (ELLA-PCT) enabling quantitative testing at the point of care without conventional laboratory instrumentation, and has a strong potential to decentralise diagnosis of infectious diseases, sup...
Y. Cai, J. Flauzino, A. Şanli et al.· medRxiv· 0 citations
Investigation of possible interference this agent may induce with typical clinical diagnostic assays given to incoming and hospitalized patients found no interference from relmapirazin was detected in the suite of metabolic and cardiac assays evaluated.
R. Dorshow, N. Morrison, Margot Borgel· International journal of tox...· 0 citations