Dual-Plane Shack-Hartmann Wavefront Sensing in a Computer Eye Model Implanted With Rotationally Asymmetric Bifocal Intraocular Lenses.

Aug 2026 · Journal of refractive surgery · Vol 42 8, pp. e759-e769 · 0 citations · 17 references
Medicine

TL;DR

By providing precise postoperative wavefronts and through-focus visual quality metrics, this approach offers a robust tool for clinical evaluation of patients implanted with these IOLs, and overcomes the phase asymmetry of segmented bifocal IOLs.

Abstract

Purpose

To evaluate wavefront aberrations in eyes implanted with rotationally asymmetric refractive bifocal intraocular lenses (IOLs) using a modified Shack-Hartmann wavefront sensor (SHWS) designed to mitigate artifactual coma.

Methods

In this simulation study, postoperative optical quality was evaluated using wavefront analysis in the presence of various ocular aberrations. A bifocal phase map (3.00 diopter [D] near addition, 50% energy distribution) was introduced in a computer eye model (5-mm pupil; 850-nm light source). The emerging wavefront displaced the simulated SHWS lenslet images from their reference positions. Wavefronts were evaluated using: (1) the standard integration method, considering lenslet images across the entire circular pupil, and (2) the proposed method, which captures SHWS data at near and far foci and combines in-focus lenslet images from the corresponding zones of the bifocal IOL. The interaction between the pupillary phase and induced wavefront aberrations (0.25 µm amplitude) was further analyzed using both methods.

Results

In the absence of ocular aberrations, standard integration detects a substantial artifactual primary vertical coma with a wavefront root mean square (RMS) of 0.57 µm. In contrast, the proposed method reduced the residual RMS to 0.01 µm. In the presence of aberrations up to the fourth order, the proposed method achieved a mean residual RMS of 35 nm, enabling diffraction-limited wavefront sensing.

Conclusions

The proposed SHWS integration method overcomes the phase asymmetry of segmented bifocal IOLs. By providing precise postoperative wavefronts and through-focus visual quality metrics, this approach offers a robust tool for clinical evaluation of patients implanted with these IOLs.

View source

Similar papers

Aug 2026

From Earlier to Updated Quadrifocal Intraocular Lens Designs: An Objective Optical Metrology Analysis.

PURPOSE To compare the optical performance of the Clareon PanOptix and Clareon PanOptix Pro intraocular lenses (IOLs) using interferometric metrology and identify measurable optical differences. SETTING Physiological Optics Laboratory, Daegu, Republic of Korea. DESIGN Experimental optical-bench investigation. MET...

E. Jeong, S. Bang, Seth M. Pantanelli et al. · 0 citations
Open access 2026

Enhanced reconstruction of diffractive intraocular lens profiles via Gabor-based holographic wavelength multiplexing

Multifocal intraocular lenses (MIOLs) have become a widely adopted solution for providing functional vision at multiple focal distances in presbyopic patients undergoing cataract surgery. As a result, reliable tools for the analysis and characterization of these lenses are of considerable interest in the ophthalmic ind...

Rosa Vila-Andrés, Anabel Martínez-Espert, J. Esteve-Taboada et al. · 0 citations
Open access Sep 2026

Effect of fogging lenses on accommodation using an open-view portable wavefront autorefractor

Purpose To investigate the effect of a + 2.00 D fogging lens on accommodative behavior during measurements with an open-view handheld wavefront autorefractor, considering refractive status and fogging duration. Methods Participants underwent six consecutive 10-second measurements per eye using an open-view autorefracto...

Pablo Bargalló, Gonzalo Guerrero, Paula Egido et al. · 0 citations
Conference Open access 2026

Wavefront diagnosis of a Fourier light-field microscope using aberrations in Zemax

Fourier light-field microscopy (FLFM) enables single-shot 3D imaging by placing a microlens array at the Fourier plane conjugate to the objective's aperture stop, yielding a spatially invariant 3D point-spread function and artifact-free reconstruction. This performance depends critically on a near-diffraction-limited w...

Utku Yerlikaya, Johnson Singh, Humeyra Caglayan · 0 citations
Open access Sep 2026

Diffractive intraocular lenses reshape 577-nm retinal laser photocoagulation spots in a model eye

To evaluate whether diffractive intraocular lenses (IOLs) reshape 577-nm retinal laser photocoagulation spots in a model eye. A standardized model eye received 577-nm continuous-wave laser photocoagulation through a retinal contact lens at fixed settings: 200-µm spot, 100 mW, and 0.2 s. One monofocal IOL (ZCB00) and tw...

Ke Wei, Feng-Tao Ji, Yong-Rong Li et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.