High-dose-rate prostate brachytherapy imaging optimization: A motorized sagittal three-dimensional ultrasound acquisition system.

Sep 2026 · Brachytherapy · 0 citations · 24 references
Medicine

TL;DR

The motorized attachment for SR3D acquisition reduced image variability and improved image quality, and may encourage broader use of sagittal reconstruction in HDR-BT, potentially improving patient outcomes.

Abstract

Background

High-dose-rate prostate brachytherapy (HDR-BT) relies on precise three-dimensional (3D) ultrasound imaging for treatment planning. Sagittally reconstructed 3D (SR3D) ultrasound images have been shown to reduce dosimetric uncertainty compared to axial approaches but are not broadly implemented due in part to image quality degradation and variability caused by non-uniform manual probe rotation during acquisition.

Purpose

We aimed to develop a system enabling motorized SR3D ultrasound acquisitions with the CIVCO EX3 stepper and evaluate improvements in image quality and scan variability.

Methods

The system was developed, allowing for automated rotation of a housed transrectal ultrasound probe in the CIVCO EX3 stepper. Image quality was assessed by comparing motorized and manual SR3D acquisitions of a wire phantom against an ideal clinical reference (slow manual scan) using peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and multi-scale SSIM (MS-SSIM) using a variety of scan speeds (2-27.5°/s). Additionally, quantitative metrics such as axial and elevational resolution and distance errors, as well as volume errors were compared between the two acquisition methods across two operators.

Results

Motorized scans were of greater similarity to an ideal reference when compared to manual acquisition regardless of acquisition speed. Manual scans showed significantly higher variability for elevational distance errors and volume errors. Additionally, manual scans had significant inter-operator variability which was not present in the motorized acquisitions.

Conclusions

The motorized attachment for SR3D acquisition reduced image variability and improved image quality. Through these improvements, this system may encourage broader use of sagittal reconstruction in HDR-BT, potentially improving patient outcomes.

View source

Similar papers

Open access Sep 2026

Dosimetric Impact of Deformable Image Registration versus Simple Summation in Multi-Fraction HDR Brachytherapy for Cervical Cancer under Standard Protocol

Objective: In high-dose-rate brachytherapy for cervical cancer, cumulative dose is commonly estimated using simple summation (SS) of dose–volume histogram parameters, assuming high-dose regions remain spatially consistent across fractions. Deformable image registration (DIR) offers an alternative approach that accounts...

Chitchaya Suwanraksa, Bheampos Rojkitjanuruk, Tanapat Suwanpan et al. · 0 citations
Open access Oct 2026

Evaluation of registration accuracy using short-arc cone-beam computed tomography for image-guided radiotherapy in breast cancer.

BACKGROUND Image-guided radiotherapy (IGRT) using cone-beam computed tomography (CBCT) enables highly accurate imaging registration; however, increased radiation exposure remains a concern. Recently, short-arc CBCT (SA-CBCT), which reduces projection data and reconstructs images, has become available. Although previous...

Shuhei Wada, M. Tominaga, M. Sasaki et al. · 0 citations
Open access Sep 2026

Clinical impact of setup shift application on image registration for hypofractionated prostate cancer treatment

Daily CBCT image-guided couch corrections maintain dosimetric accuracy during hypofractionated prostate radiotherapy and support safe implementation of tight PTV margins in resource-constrained public-sector healthcare settings.

Leanore Legoale-Sebopa, Sheynaz Bassa, H. V. Maselesele · 0 citations
Open access Sep 2026

Feasibility of quantifying intra-fractional patient positional error for head and neck IMRT using radiomic features of EPID-measured fluence maps: a phantom study

This study aimed to evaluate the feasibility of estimating patient positional errors during head-and-neck fixed-field intensity-modulated radiation therapy (IMRT) using a machine-learning (ML) regression models based on radiomics features of patient-transmitted fluence maps acquired by an electronic portal imaging devi...

Takayuki Nishikata, S. Utsunomiya, Madoka Sakai et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.