Skip to content
Open access

Bringing usability testing to the front lines: IV smart pump evaluation with critical care nurses at a national conference

Sep 2026 · MethodsX · Vol 17, pp. 104140 · 0 citations · 29 references
Medicine

TL;DR

The method brought a standardized simulation environment to a national critical-care nursing conference while preserving experimental control through consistent pump stations, scripted clinical scenarios, counterbalanced device order, and common data-collection procedures.

Abstract

Intravenous smart pump (IVSP) usability evaluations are often conducted in single-site laboratory or simulation settings, which can limit access to experienced frontline nurses. This methods article describes a field-deployable, conference-based approach for evaluating IVSP usability with practicing critical-care nurses. The method brought a standardized simulation environment to a national critical-care nursing conference while preserving experimental control through consistent pump stations, scripted clinical scenarios, counterbalanced device order, and common data-collection procedures. The approach integrates direct observation of programming performance with wearable eye tracking, physiologic monitoring, subjective workload ratings, healthcare-specific usability assessment, think-aloud comments, and device preference ranking. The method is intended for researchers, health systems, manufacturers, and nurse-engineer teams seeking a scalable way to evaluate bedside technologies with intended users. • Adapts traditional laboratory-based usability testing into a portable, conference-based method. • Combines standardized clinical tasks with multimodal measures of visual attention, workload, usability, and performance. • Provides a reproducible framework for evaluating medical-device usability with frontline clinicians.

Read PDF

Similar papers

Open access Aug 2026

Lessons from the field: usability (engineering) in regulated healthcare

It is argued that usability engineering should be a formative, safety-critical discipline integrated throughout the product lifecycle, including aftermarket deployment, and not merely a compliance task.

Preetha Moorthy, Till Nagel, Eva Hornecker · 0 citations
Review Apr 2026

Human-Centered Design and Usability Evaluation of an EHR-Integrated Clinical Decision Support Prototype for Pediatric Surgical Site Infection Risk

Abstract Objectives This study aimed to design and evaluate the usability of an electronic health record (EHR)-integrated clinical decision support prototype supporting pediatric surgical site infection (SSI) risk estimation and preoperative optimization using a multiphase human-centered design approach. Methods We con...

C. Chan, Karthik Balakrishnan, Aaron W. Scheffler et al. · 0 citations
Review Open access Sep 2026

A Transferable Evaluation Framework for Mobile Health Data Collection in Shift-Work Nurses: Development and Feasibility Study

Abstract Background Shift-work nurses experience substantial variability in work schedules, health behaviors, and recovery patterns, which traditional retrospective surveys fail to capture due to recall bias. Although mobile health (mHealth) tools offer ecological momentary data capture, existing off-the-shelf survey p...

Kyulhee Park, K. Han · 0 citations
Review Open access Aug 2026

Usability and user experience of an educational electronic prescribing system to train medical students: a mixed-methods approach

Prescribing is a core responsibility of physicians, with electronic (e-)prescribing becoming the global standard. However, e-prescribing is insufficiently integrated in medical curricula, creating a gap between education and actual practice. To address this, we developed Edups, an educational electronic prescribi...

Samir El Abdouni, L. Kalfsvel, Tolunay Görcu et al. · 0 citations
Open access Sep 2026

Usability and Quality of UCISAB, a Clinical Decision Support System for Optimization of Analgesia and Sedation Infusions in the ICU: A Multicenter, Cross-Sectional, Mixed-Methods Usability Evaluation in Colombia

UCISAB showed favorable perceived quality and a high stated willingness to use, while usability remained marginal; further interface and pharmacological refinements, together with prospective clinical validation in real-world settings, are required before routine implementation.

German David Castillo-Suárez, Manuel Mena, N. Proaños et al. · 0 citations
Open access Jan 2026

Perceived Nursing Interruptions During Medication Administration: Development and Validation of a Questionnaire Among Chinese Nurses

The Q‐NIMA demonstrates robust psychometric properties, including high reliability, validity, and stability, and serves as a valuable tool to evaluate how nurses perceive nursing interruptions during medication administration.

Xiao-Qian Dong, Na Chen, Cui-Lan Dong et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.