This work validates a hybrid spectral CT system, comprised of clinical-grade components, acquiring four distinct effective spectra and applying efficient noise-reducing weighting schemes to compare iodine noise and bias against conventional kVp-Switching (kVp-S).
Abstract
Spectral computed tomography (CT) is a burgeoning quantitative imaging technique with applications in oncologic diagnostics, prognostic prediction, tissue perfusion studies, and treatment follow-up. While normalized iodine concentration values have been correlated with microenvironmental biophysical changes, obtaining accurate iodine concentrations, particularly at low concentrations remains difficult due to varying spectral CT instrumentation performance. Hybrid spectral CT systems, combining multiple spectral CT instrumentation techniques, address these quantitation insufficiencies by increasing spectral separation but have not been evaluated on a clinically analogous platform. We validate a hybrid spectral CT system, comprised of clinical-grade components, acquiring four distinct effective spectra and applying efficient noise-reducing weighting schemes to compare iodine noise and bias against conventional kVp-Switching (kVp-S). Two tube current levels (50, 350 mA) and three duty cycle ratios (33/67, 50/50, 75/25) were implemented to elucidate radiation dose exposure and kVp-S parameterization impact. A standard quality assurance (QA) and patient-derived, abdominal IodinePrint phantom were scanned on the system. The average absolute bias in iodine density images of the QA phantom was comparable across acquisition techniques, below 0.5 mg/mL, while quantitative noise improved by 22% using noise-optimized weighting schemes. In the IodinePrint phantom aorta and pancreas structures, the noise-optimized weighting scheme increased signal-to-noise ratio (SNR) by 1.3x compared to kVp-S alone. These results highlight the increased precision of hybrid, multi-channel spectral CT systems and motivate CT designs that enable robust CT biomarker development.
In PET/CT imaging, CT frequently accounts for more than 50% of the total radiation exposure, even when used solely for attenuation correction and anatomical localisation. This poses a major limitation for dose-sensitive populations, including pregnant women and paediatric patients. This study systematically evalu...
Georgios I. Angelis, Y. Gholami, K. Willowson et al.· EJNMMI Physics· 0 citations
The findings support the use of optimized low dose CT protocols in SPECT/CT without compromising quantitative reliability, helping radiographers and nuclear medicine technologist balance radiation protection with diagnostic requirements.
E. Karić, M. Kukuljan, L. Lezaic et al.· Radiography· 0 citations
The development of Long Axial Field-of-View positron emission tomography combined with computed tomography warrants reconsideration of the administered activity required for clinical imaging. This study investigated the relationship between body size, administered activity (18F), and image quality (IQ) using two patien...
Nanna Overbeck, Zuzanna Czeluśniak, Søren Holm et al.· Physics in Medicine and Biol...· 0 citations
Abstract Background Technological developments in computed tomography (CT) have increased the diversity of acquisition and reconstruction strategies available for clinical imaging. Task‐based image quality metrics that account for patient‐size variability are needed to characterize reconstructed image performance and s...
Isabelle Fitton, J. Greffier, D. Dabli et al.· Medical Physics (Lancaster)· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.