Accuracy of implant positioning in total ankle arthroplasty using patient-specific instrumentation: A weightbearing CT analysis of landmark-based parameters.
Sep 2026· Foot and Ankle Surgery· 0 citations· 26 references
Medicine
TL;DR
InBONE Prophecy PSI accurately reproduced planned distance-based positioning, while sagittal talar alignment was more variable.
Abstract
Background
Patient-specific instrumentation (PSI) may improve implant positioning in total ankle arthroplasty (TAA), but three-dimensional validation with postoperative weightbearing CT (WBCT) remains limited. This study assessed the accuracy of the INBONE Prophecy system.
Methods
Fifty-four patients undergoing INBONE II TAA with available Prophecy planning and postoperative WBCT were retrospectively reviewed. Planned and postoperative implant positions were compared for medial tibial component-to-medial malleolus distance, talar stem-to-subtalar joint distance, and talar cut relative to the horizontal axis. Reliability was evaluated using intraclass correlation coefficients.
Results
Distance measurements showed high accuracy, with most cases within 1 mm. Poor accuracy (>2 mm) occurred in 20% of medial malleolar and 7.7% of subtalar measurements. ICCs ranged from 0.79 to 0.95. The mean absolute angular difference for talar cut was 2.91°.
Conclusion
INBONE Prophecy PSI accurately reproduced planned distance-based positioning, while sagittal talar alignment was more variable.
Unicompartmental knee arthroplasty (UKA) is an effective treatment for medial compartment knee osteoarthritis, with its outcomes highly dependent on the accuracy of prosthesis positioning. Conventional UKA, performed using traditional instrumentation, often relies heavily on surgical experience and may lead to variabil...
Kun Zhang, Jin-Hua Zhang, Shang Gao et al.· 3D Printing in Medicine· 0 citations
In this laboratory-based TAA study, it is found that both CAN and PSI demonstrated advantages over SRG instrumentation regarding implant placement accuracy and the observed reduction in implant placement variability has potential positive implications for surgeons.
Cole Herbel, Ramiro Lopez, Matthew C. Rueff et al.· Journal of the American Acad...· 0 citations
Transfibular TAA is capable of correcting severe coronal tibiotalar malalignment in the primary treatment of end-stage ankle arthritis and demonstrated postoperative patient-reported and radiographic outcomes similar to those with neutral ankle alignment at midterm follow-up, without an increased risk of reoperations.
L. Schon, Jonathan Day, Morgan Motsay et al.· JB & JS open access· 0 citations
BACKGROUND
The aim of this study was to evaluate whether preoperative medial proximal tibial angle (MPTA) influences clinical outcomes following mobile-bearing medial unicompartmental knee arthroplasty (UKA).
METHODS
This retrospective study included 454 consecutive patients who had antero-medial osteoarthritis and u...
Tomofumi Kinoshita, K. I. Bunyoz, Kristian R. L. Mortensen et al.· Journal of Arthroplasty· 0 citations
TKA performed with domestic cementless knee prostheses featuring 3D-printed trabecular metal interface shows no obvious disadvantages compared with cemented prostheses in terms of postoperative function, complication rate, and prosthesis survival rate, which represents a safe and viable prosthesis option.
Hao Chai, Yong-Qiang Sun, Xiao-Bo Chen et al.· Zhongguo xiu fu chong jian w...· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.