Aug 2026· PLoS ONE· Vol 21, pp. e0356995 - e0356995· 0 citations· 22 references
Medicine
TL;DR
The three devices showed acceptable agreement at the population level, and notably, the average magnitude of disagreement was limited, at the individual level, however, differences exceeded clinically relevant thresholds in a meaningful proportion of patients, including some toric IOL candidates.
Abstract
Accurate keratometry is essential for toric intraocular lens (IOL) planning, yet most prior studies evaluated inter-device agreement under standardized research conditions that may not reflect everyday practice. This retrospective study evaluated keratometry agreement among the KR-800, IOLMaster 500, and Pentacam HR under real-world clinical conditions in 227 eyes of cataract surgery candidates. Corneal astigmatism was converted to power vector components: J0 (with/against-the-rule astigmatism) and J45 (oblique astigmatism). Population-level agreement was assessed using linear mixed-effects models and intraclass correlation coefficients (ICC), while individual-level agreement was evaluated using Bland-Altman analysis. Agreement in toric IOL candidacy and the predicted residual astigmatism arising from inter-device differences were also examined. No significant population-level differences were observed among devices (J0: p = 0.138; J45: p = 0.233), with ICC values of 0.894 for J0 and 0.748 for J45, respectively. However, the 95% limits of agreement were wide (0.80–1.21 D). When any device pair was considered, the predicted residual astigmatism exceeded 0.50 D in 64.8% of patients, and the Pentacam HR classified fewer eyes as toric IOL candidates than the reflection-based devices. Among toric IOL candidates (≥1.0 D astigmatism), 21–33% demonstrated axis differences exceeding 10°, a threshold associated with clinically relevant reduction in astigmatic correction. Axis differences were inversely correlated with astigmatism magnitude (Spearman’s rho = −0.220 to −0.272, all p-values < .001). The three devices showed acceptable agreement at the population level, and notably, the average magnitude of disagreement was limited. At the individual level, however, differences exceeded clinically relevant thresholds in a meaningful proportion of patients, including some toric IOL candidates. Using a single device consistently throughout surgical planning may be prudent, although the clinical impact of these differences was not directly assessed and warrants prospective validation.
Introduction: Preoperative biometry plays a pivotal role in predicting visual outcomes after cataract surgery. Corneal astigmatism can significantly influence uncorrected visual acuity (UCVA). In clinical settings where only non-toric intraocular lenses (IOLs) are available, understanding the impact of astigmatism beco...
Milzan Murtadha, Sagung Gede Indrawati, Amanda Nur Shinta Pertiwi· Ophthalmologica Indonesiana· 0 citations
PURPOSE
To compare the repeatability of and describe the agreement between the IOLMaster 700 (Carl Zeiss Meditec AG) and Eyestar 900 (Haag-Streit) under identical test conditions.
METHODS
This was a comparative cross-sectional study conducted at a single-surgeon ophthalmology clinic in Australia. Adults presenting fo...
Malak Habib, B. Dyer, Shu-Kay Ng et al.· Journal of refractive surger...· 0 citations
Background/Objectives: Accurate preoperative biometry is essential for refractive cataract surgery. Pentacam AXL WAVE is an emerging multimodal platform integrating ocular biometry, anterior segment imaging, and corneal assessment, yet its interchangeability with established devices remains unclear. This study aimed to...
Rong Zhao, Zhi-Yang Zhang, Su-Yu Wang et al.· Journal of Clinical Medicine· 0 citations
Purpose: This study aimed to compare the repeatability and agreement of corneal tomographic measurements between the Anterion Cornea App and Pentacam HR in keratoconus (KCN). Methods: Eyes with KCN were imaged three consecutive times with each device. Intra-device repeatability was assessed using intraclass correlation...
Ramadhan Ahmed, K. Reddy, Mahad Mohamed et al.· Bioengineering· 0 citations