Dosimetric Evaluation of MR-Guided IMRT and VMAT Techniques for Early-Stage Glottic Cancer: A Pilot Comparative Planning Study

Aug 2026 · Medicina · Vol 62, pp. 1507 · 0 citations · 26 references
Medicine

TL;DR

MR-guided IMRT provided superior dose conformity while maintaining target coverage comparable to TrueBeam STx with clinically acceptable OAR sparing, although selected laryngeal structures and the esophagus differed significantly.

Abstract

Background and Objectives: Early-stage glottic cancer requires highly precise radiotherapy due to laryngeal motion and the proximity of critical structures. Magnetic resonance-guided radiotherapy (MRgRT) enables real-time visualization and motion management, potentially improving dose conformity while reducing treatment margins. This pilot comparative planning study compared the dosimetric characteristics of MR-guided intensity-modulated radiotherapy (IMRT) delivered on the Unity MR-linac with volumetric modulated arc therapy (VMAT) plans generated for the TrueBeam STx and Halcyon systems. Materials and Methods: Ten patients with T1–T2N0 glottic squamous cell carcinoma were included. For each patient, treatment plans prescribing 42.5 Gy in five fractions were generated using all three systems. Planning target volume (PTV) coverage, conformity index (CI), homogeneity index (HI), organ-at-risk (OAR) doses, monitor units, and beam-on times were compared using Friedman and post hoc Wilcoxon signed-rank tests. Results: Unity achieved superior dose conformity (CI 0.97 ± 0.01 vs. 0.78 ± 0.15 for TrueBeam STx and 0.58 ± 0.19 for Halcyon; p < 0.001) and the highest mean dose, while PTV coverage (D95% and D98%) was comparable to TrueBeam STx and higher than Halcyon. Reduced treatment margins resulted in significantly smaller PTV volumes (11.09 ± 3.87 cm3 vs. 17.43 ± 4.75 cm3; p = 0.0001). Halcyon demonstrated the most homogeneous dose distribution (HI 7.88 ± 3.14), whereas TrueBeam STx showed intermediate homogeneity (HI 12.81 ± 3.90). Most OAR dose metrics were comparable among techniques, although selected laryngeal structures and the esophagus differed significantly. Beam-on time was longest for Unity. Conclusions: MR-guided IMRT provided superior dose conformity while maintaining target coverage comparable to TrueBeam STx with clinically acceptable OAR sparing. Prospective clinical studies are required to determine whether these dosimetric advantages translate into improved patient outcomes.

Read PDF

Similar papers

Open access Oct 2026

Evaluating plan quality using field-in-field, static and dynamic intensity-modulated radiotherapy techniques in pancreatic cancer: A dosimetric study

Objective: The present study aims to compare field-in-field (FIF) radiation therapy, static intensity-modulated radiation therapy (S-IMRT), and dynamic static intensity-modulated radiation therapy (D-IMRT) techniques concerning dosimetric benefits for the treatment of pancreatic cancers. Materials and Methods: Sixteen...

G. Inan, Hamit Başaran, Osman Vefa Gül · 0 citations
Aug 2026

MRI-Guided Reduced-Volume RT for Early-Stage Glottic Cancer: A Prospective Phase 1 Trial with Quantification of Laryngeal Motion.

PURPOSE Five to 7 weeks of whole-larynx radiation (RT) is standard for treatment of early-stage glottic cancer. Previous studies evaluated CT-guided hypofractionated/SBRT treatment to a smaller volume. Magnetic resonance-guided radiotherapy (MRgRT) allows real-time motion tracking and gating by continuously imaging the...

J. Leeman, A. Droznin, C. D. Dwyer et al. · 0 citations
Open access Sep 2026

MR-guided ultra-hypofractionated radiotherapy for early-stage glottic cancer: A phase I feasibility study

Introduction Ultra-hypofractionation for early-stage glottic cancer reduces the number of hospital visits, but requires precise delivery to minimize the risk of radiation-induced toxicity. Magnetic Resonance (MR)-guided radiotherapy is a delivery technique that enables visualization of the tumor and laryngeal cartilage...

S. M. L. Linden, M. Philippens, J. Rijken et al. · 0 citations
Open access Sep 2026

Forward-Planned Multisegmented Technique in Resource-Limited Settings: A Feasible Dosimetric Upgrade for 3D-CRT in Locally Advanced Nasopharyngeal Carcinoma

In resource-limited settings, intensity-modulated radiotherapy (IMRT) is often unavailable, forcing reliance on three-dimensional conformal radiotherapy (3D-CRT) for advanced nasopharyngeal carcinoma (NPC). This limits curative potential due to challenges in sparing organs-at-risk (OARs) while covering complex target v...

AS Badiane, M. Dieng, Lef Edimo et al. · 0 citations
Sep 2026

Comprehensive and Comparative Dosimetric Characterization of VMAT and IMRT in Brain Cancer Treatment Planning: Target Coverage, OAR Sparing, Monitor Units, Dose Distribution, Low-Dose Volumes, and Delivery Efficiency

Purpose: A detailed dosimetric comparison is conducted between VMAT and IMRT modalities for brain tumor treatment plans concerning prescription doses of 60, 66, and 70 Gy. Material and Methods: A total number of forty-three brain tumor patients were selected by means of a retrospective study in the Radiotherapy Depart...

Abbood Abbas Abbood · 0 citations