Sep 2026· Physics in Medicine and Biology· Vol 71· 0 citations· 49 references
PhysicsMedicine
TL;DR
This work establishes a robust, cross-validated computational baseline for brachytherapy dosimetry using general-purpose MC codes under realistic photon emission spectra under realistic, unsimplified photon emission spectra.
Abstract
Objective. This study presents a rigorous, simultaneous Monte Carlo (MC) intercomparison of three general-purpose computational codes, MCNP6.2, TOPAS 3.9, and PHITS 3.341, for the calculation and validation of TG-43 dosimetric parameters for the clinical microSelectron-HDR v2 192Ir and Theragenics-AgX100 125I brachytherapy sources under realistic, unsimplified photon emission spectra. Approach. Simulations were performed in a spherical liquid water phantom with a radius of 18.5 cm to ensure full-scatter conditions. Dose was scored in small, distance-dependent spherical water scoring cells with varying radii. Discrete, unsimplified photon emission spectra from ICRP Publication 107 were implemented in all three codes. To address gaps in the literature, a computationally efficient simplified geometry previously adopted for the 192Ir source was evaluated, while the 125I source was modeled with high geometric fidelity. The radial dose and anisotropy functions were calculated, the radial dose data were fitted using fifth-degree polynomials, and the resulting datasets were benchmarked against established TG-43 consensus reference data. Main results. The three MC codes demonstrated good overall agreement. For the 192Ir source, the simplified geometry reproduced the consensus reference data within the ±2% criterion over the clinically relevant radial range, demonstrating its dosimetric suitability within the TG-43 framework. For the low-energy 125I source, larger deviations, reaching approximately 21%, were observed at short radial distances ( r ⩽ 0.5 cm), consistent with the greater sensitivity of low-energy photon transport to attenuation and shielding effects associated with the source encapsulation and internal components. At distances beyond approximately 1.0 cm, the results showed substantially improved agreement, with deviations generally remaining within ±3%. Fifth-degree polynomial fitting accurately reproduced the radial dose function, with deviations predominantly remaining below the recommended 2% criterion within the clinically relevant radial intervals. Significance. This work establishes a robust, cross-validated computational baseline for brachytherapy dosimetry using general-purpose MC codes under realistic photon emission spectra. The validation of the simplified 192Ir geometry supports its use in computationally efficient simulations, while the characterization of the low-energy 125I source provides relevant information for high-precision dosimetry and future investigations of low-energy radiation interactions in radiosensitization studies.
INTRODUCTION
Numerous dosimetric studies have reported clinically insignificant differences between iridium-192 and cobalt-60. However, such comparison studies were conducted using TG-43 algorithms where heterogeneity is not accounted. Model based dose calculation algorithms (MBDCA) significantly improve dosimetric acc...
Background: This study aimed to evaluate the feasibility and dosimetric characteristics of 90Sr/90Y and 188Re sources for treating superficial oral rhabdomyosarcoma (RMS) using a surface mold brachytherapy approach.
Materials and Methods: In this Monte Carlo simulation study, TG-149 dosimetric parameters were calculat...
Shahriyar Seraj, Shiva Rahbar Yazdi, Soroush Najafi Sarvestani et al.· Iranian Journal of Pediatric...· 0 citations
The accuracy of TG-43 algorithm against Monte Carlo (MC) simulation is investigated, confirming its adequacy in homogeneous phantom, and the necessity of adopting Model-based calculation algorithms such as MC to improve the dosimetric precision and treatment accuracy in HDR brachytherapy is emphasized.
R. Shende, S. J. Dhoble, Yassir El Ghazi· Applied Radiation and Isotop...· 0 citations
Permanently implanted iodine-125 (125I) brachytherapy sources are a radiological technique used to treat localized cancers. The TG-43U1 group of the AAPM emphasizes the importance of calculating the dose distribution around brachytherapy seeds before their clinical use or approval for therapeutic purposes. In this work...
Manal El Asri, O. El Bounagui· Health Physics· 0 citations
An initial 3D electronic detector array prototype for radiotherapy, consisting of 2645 diodes (8.0 mm pitch), was modelled with Geant4 and MCNP6.2 Monte Carlo codes to simulate irradiations of 2×2, 3×3, 4×4, and 5×5 cm2 field sizes with a 6 MV photon beam. Comparisons with Treatment Planning Software (TPS) data reveale...
R. Ocampo, D. Roa, Enzo Aucca et al.· Journal of Physics, Conferen...· 0 citations
Stereotactic radiosurgery delivers an ablative single-fraction dose to small intracranial targets, and the reported dose depends on the calculation algorithm. This exploratory case series quantified the change in reported dosimetry when plans clinically finalised with the anisotropic analytical algorithm (AAA) were rec...
K. Utitsarn, Thepphithak Watthanasarn, Wirasinee Chaloemchawalit et al.· Biomedical engineering and p...· 0 citations