The new MS-39 device demonstrated moderate-to-good precision in corneal epithelial aberration measurements for FFKC, mild, moderate and total KC, and the potential of the device for early diagnosis and monitoring was still evident.
Abstract
To establish a precision reference benchmark for corneal epithelial aberration measurements across different stages of keratoconus (KC) by evaluating the precision of a new spectral-domain optical coherence tomography–Placido disk device (MS-39; CSO). Eyes with KC were classified into forme fruste keratoconus (FFKC), mild, moderate, and severe disease stage. Three consecutive corneal epithelial aberration measurements were obtained with the MS-39 by two skilled operators. The intraobserver repeatability and interobserver reproducibility were evaluated using within-subject standard deviation (Sw), test–retest variability (TRT), and intraclass correlation coefficient (ICC). The prospective study enrolled 141 eyes from 141 KC subjects. Subgroups included 32 eyes with FFKC, 29 with mild, 45 with moderate, and 35 with severe KC. In the FFKC, mild, moderate, and total groups, Sw and TRT for intraobserver repeatability were ≤ 0.18 μm and ≤ 0.50 μm, respectively. ICCs were ≥ 0.884 for most parameters except trefoil and astigmatism II. The Sw and TRT for interobserver reproducibility were ≤ 0.15 μm and ≤ 0.41 μm, respectively. ICCs were ≥ 0.917 except for trefoil and astigmatism II. In the severe group, the ICCs for intraobserver repeatability and interobserver reproducibility were 0.366 to 0.904 and 0.493 to 0.908. The new MS-39 device demonstrated moderate-to-good precision in corneal epithelial aberration measurements for FFKC, mild, moderate and total KC. Despite the decrease associated with increasing KC severity, the potential of the device for early diagnosis and monitoring was still evident. The instrument’s precision should be tailored to disease severity. Stage-specific TRT thresholds for epithelial aberration measurements were proposed for KC management.
Objectives: To assess the repeatability and reliability of anterior and posterior corneal measurements in keratoconus patients who underwent corneal allogenic intrastromal ring segment (CAIRS) implantation using a Scheimpflug-based tomography system. Methods: This prospective observational study included 42 eyes: 20 ke...
Uğur Tunç, Dilan Çolak, Burcu Yakut et al.· Beyoglu Eye Journal· 0 citations
Purpose: This study aimed to compare the repeatability and agreement of corneal tomographic measurements between the Anterion Cornea App and Pentacam HR in keratoconus (KCN). Methods: Eyes with KCN were imaged three consecutive times with each device. Intra-device repeatability was assessed using intraclass correlation...
Ramadhan Ahmed, K. Reddy, Mahad Mohamed et al.· Bioengineering· 0 citations
To compare corneal epithelial thickness (CET) profiles among different types of regular astigmatism and early keratoconus (KCN) using the MS-39 anterior segment optical coherence tomography (AS-OCT) and to identify disease-specific epithelial asymmetry patterns, regional epithelial thickness, and ROC performance. This...
PURPOSE
To evaluate the repeatability of anterior segment parameters measured with a novel swept source optical coherence tomography (SS-OCT) instrument, and to compare these anterior segment parameters with those measured with the CASIA2.
METHODS
This prospective case series study included 119 eyes from 119 particip...
Zheng-Yuan Qu, Yan Huo, Rui-Si Xie et al.· Photodiagnosis and Photodyna...· 0 citations
PURPOSE
To compare the repeatability of and describe the agreement between the IOLMaster 700 (Carl Zeiss Meditec AG) and Eyestar 900 (Haag-Streit) under identical test conditions.
METHODS
This was a comparative cross-sectional study conducted at a single-surgeon ophthalmology clinic in Australia. Adults presenting fo...
Malak Habib, B. Dyer, Shu-Kay Ng et al.· Journal of refractive surger...· 0 citations