Establishing a precision reference benchmark for corneal epithelial aberration measurements using a new hybrid device across keratoconus stages

Sep 2026 · Eye and Vision · Vol 13 · 0 citations · 33 references
Medicine

TL;DR

The new MS-39 device demonstrated moderate-to-good precision in corneal epithelial aberration measurements for FFKC, mild, moderate and total KC, and the potential of the device for early diagnosis and monitoring was still evident.

Abstract

To establish a precision reference benchmark for corneal epithelial aberration measurements across different stages of keratoconus (KC) by evaluating the precision of a new spectral-domain optical coherence tomography–Placido disk device (MS-39; CSO). Eyes with KC were classified into forme fruste keratoconus (FFKC), mild, moderate, and severe disease stage. Three consecutive corneal epithelial aberration measurements were obtained with the MS-39 by two skilled operators. The intraobserver repeatability and interobserver reproducibility were evaluated using within-subject standard deviation (Sw), test–retest variability (TRT), and intraclass correlation coefficient (ICC). The prospective study enrolled 141 eyes from 141 KC subjects. Subgroups included 32 eyes with FFKC, 29 with mild, 45 with moderate, and 35 with severe KC. In the FFKC, mild, moderate, and total groups, Sw and TRT for intraobserver repeatability were ≤ 0.18 μm and ≤ 0.50 μm, respectively. ICCs were ≥ 0.884 for most parameters except trefoil and astigmatism II. The Sw and TRT for interobserver reproducibility were ≤ 0.15 μm and ≤ 0.41 μm, respectively. ICCs were ≥ 0.917 except for trefoil and astigmatism II. In the severe group, the ICCs for intraobserver repeatability and interobserver reproducibility were 0.366 to 0.904 and 0.493 to 0.908. The new MS-39 device demonstrated moderate-to-good precision in corneal epithelial aberration measurements for FFKC, mild, moderate and total KC. Despite the decrease associated with increasing KC severity, the potential of the device for early diagnosis and monitoring was still evident. The instrument’s precision should be tailored to disease severity. Stage-specific TRT thresholds for epithelial aberration measurements were proposed for KC management.

Read PDF

Similar papers

Open access 2026

Repeatability and Reliability of Corneal Tomography in Keratoconus Patients with Corneal Allogenic Intrastromal Ring Segments

Objectives: To assess the repeatability and reliability of anterior and posterior corneal measurements in keratoconus patients who underwent corneal allogenic intrastromal ring segment (CAIRS) implantation using a Scheimpflug-based tomography system. Methods: This prospective observational study included 42 eyes: 20 ke...

Uğur Tunç, Dilan Çolak, Burcu Yakut et al. · 0 citations
Open access Sep 2026

Measurement Repeatability and Concordance of Corneal Tomography in Keratoconus Using Anterion Versus Pentacam

Purpose: This study aimed to compare the repeatability and agreement of corneal tomographic measurements between the Anterion Cornea App and Pentacam HR in keratoconus (KCN). Methods: Eyes with KCN were imaged three consecutive times with each device. Intra-device repeatability was assessed using intraclass correlation...

Ramadhan Ahmed, K. Reddy, Mahad Mohamed et al. · 0 citations
Open access Oct 2026

Regional corneal epithelial thickness and asymmetry patterns in early keratoconus versus regular astigmatism using MS-39 anterior segment OCT: a comparative cross-sectional study

To compare corneal epithelial thickness (CET) profiles among different types of regular astigmatism and early keratoconus (KCN) using the MS-39 anterior segment optical coherence tomography (AS-OCT) and to identify disease-specific epithelial asymmetry patterns, regional epithelial thickness, and ROC performance. This...

Yasaman Hadi, Farideh Kheimeh Kabood, Parya Abdolalizadeh et al. · 0 citations
Open access Aug 2026

Repeatability and Agreement between novel SS-OCT VG2 and CASIA2 in Anterior Segment Parameters.

PURPOSE To evaluate the repeatability of anterior segment parameters measured with a novel swept source optical coherence tomography (SS-OCT) instrument, and to compare these anterior segment parameters with those measured with the CASIA2. METHODS This prospective case series study included 119 eyes from 119 particip...

Zheng-Yuan Qu, Yan Huo, Rui-Si Xie et al. · 0 citations
Sep 2026

The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.

PURPOSE To compare the repeatability of and describe the agreement between the IOLMaster 700 (Carl Zeiss Meditec AG) and Eyestar 900 (Haag-Streit) under identical test conditions. METHODS This was a comparative cross-sectional study conducted at a single-surgeon ophthalmology clinic in Australia. Adults presenting fo...

Malak Habib, B. Dyer, Shu-Kay Ng et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.