Manufacturing trueness, internal fit, and planned-to-intraoral spatial concordance of monolithic zirconia crowns fabricated with two commercial ti-base systems: a clinical cross-sectional study

Aug 2026 · BMC Oral Health · 0 citations

TL;DR

Planned-to-intraoral spatial RMS values were consistently higher than manufacturing RMS values in both systems, indicating that high manufacturing trueness alone does not imply high planned-to-intraoral spatial concordance.

Abstract

The aim of this clinical cross-sectional study was to evaluate manufacturing trueness, internal fit, planned-to-intraoral spatial concordance, and proximal contact discrepancy in implant-supported monolithic zirconia crowns fabricated within two commercial ti-base restorative systems. Thirty-one patients received implant-supported single-unit monolithic zirconia crowns fabricated within either a Flat or Tab commercial ti-base restorative system using a fully digital workflow. The systems differed in manufacturer and associated components or parameters, including proprietary scan bodies and CAD libraries, abutment height, and manufacturer-recommended cement gap; accordingly, this was a system-level comparison rather than an isolated test of ti-base interface geometry. For each restoration, four STL datasets represented the initial intraoral situation, the CAD crown design, the manufactured crown, and the intraoral condition after fitting. A modified triple-scan protocol and metrology-grade software were used. Manufacturing trueness and planned-to-intraoral spatial concordance were quantified using root mean square deviation values. Because the planned and observed datasets were propagated within the same digital workflow and no independent external reference was used, the spatial outcome was interpreted as workflow concordance rather than absolute clinical trueness. Internal fit was assessed by triple-scan gap analysis, and proximal contact discrepancies were digitally measured. Statistical analyses were performed using Student’s t test and repeated-measures ANOVA with significance set at p  < 0.05. No statistically detectable between-system differences were found for manufacturing trueness, internal fit, planned-to-intraoral spatial concordance, or proximal contact discrepancies ( p  > 0.05). Spatial RMS values were consistently higher than manufacturing RMS values in both systems ( p  < 0.001). Internal fit values for both systems were within commonly cited clinical thresholds. Both commercial ti-base systems showed mean manufacturing RMS values below 60 μm and internal fit within commonly cited clinical limits. Planned-to-intraoral spatial RMS values were consistently higher than manufacturing RMS values in both systems, indicating that high manufacturing trueness alone does not imply high planned-to-intraoral spatial concordance. Manufacturing trueness alone may be insufficient to characterize planned-to-observed intraoral concordance; evaluation of the final intraoral position provides complementary workflow information.

Read PDF

Similar papers

Open access Aug 2026

The effect of build angle and intraarch connector design on the accuracy and fit of additively manufactured complete-arch implant-supported interim fixed prostheses.

Within the limitations of this single-system in vitro study, AM complete-arch implant-supported interim fixed prostheses in a 0° no-bar configuration demonstrated accuracy comparable to SM HIPC, but the marginal gaps found in some groups were beyond clinical acceptability.

Steven Sonderegger, Pedro Molinero Mourelle, Manuel Aeschbacher et al. · 0 citations
Open access Jul 2026

Accuracy of scannable anatomic healing abutments for implant-supported crown and fixed partial prostheses: An in vitro comparison with conventional scan bodies.

STATEMENT OF PROBLEM Accurate digital implant scanning is critical for the fit, function, and biological stability of implant-supported restorations. Anatomic scannable healing abutments (AHAs) have been proposed as an alternative to conventional scan bodies (SBs) to simplify digital workflows and facilitate the develo...

Paolo Nava, Carlo Nava, David Dunbar et al. · 0 citations
Aug 2026

Adjustment Efficiency and Internal Fit of Screw-Retained Zirconia Crowns on Additively Manufactured Dental Casts: An In Vitro Study on the Effect of Printing Layer Thickness and Digital Implant Analog Design.

PURPOSE To evaluate how the layer thickness of additively manufactured casts and implant analog design influence the adjustment efficiency and internal fit of screw-retained zirconia crowns, compared with those adjusted on conventional stone casts. MATERIALS AND METHODS A master mandibular model with an implant and i...

Mohamed El Gedaily, Mustafa Borga Donmez, Munir Demirel et al. · 0 citations
Aug 2026

In Vivo Assessment of Complete-Arch Implant Impressions Using a Clinical Seating Score: A Proof-of-Principle Case Report.

The CSS may potentially provide a simple and standardized chairside method for evaluating the passive seating of complete-arch implant frameworks and help clinicians identify clinically relevant misfits before delivery of the definitive prosthesis and support more predictable decision-making during framework verificati...

Francesco Mintrone, F. Ferri, Rostyslav Havryltsiv et al. · 0 citations
Open access Aug 2026

Monolithic Zirconia Versus Porcelain-Fused-to-Metal Implant-Supported Fixed Dental Prostheses in the Posterior Region: 3-Year Results of a Randomized Controlled Clinical Trial.

OBJECTIVES To compare the clinical, biological, and technical outcomes of posterior three-unit implant-supported fixed dental prostheses (iFDPs) fabricated from monolithic zirconia (MonoZR) or porcelain-fused-to-metal (PFM). MATERIALS AND METHODS In this prospective randomized controlled trial, partially edentulous p...

Aspasia Pachiou, Raphael Singh, Ronald E. Jung et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.