Operator experience on robotic and dynamic computer-assisted implant surgery (CAIS): an in vitro controlled study.

Aug 2026 · E -journal of dentistry · Vol 176, pp. 106993 · 1 citation · 21 references
Medicine

Abstract

Objective

To evaluate the influence of operator experience on implant placement accuracy and primary implant stability in dynamic and robotic computer-assisted implant surgery (d-CAIS and r-CAIS).

Materials And Methods

This controlled in vitro study evaluated three levels of operator experience: postgraduate students without previous CAIS experience (G1), implant dentistry experts without prior experience in either d-CAIS or r-CAIS (G2), and implant dentistry experts with more than 10 years of experience using both systems (G3). Coronal, apical, and angular deviations between planned and final implant positions were assessed using inspection software (Geomagic X Control, 3D Systems, Rock Hill, SC, USA). Primary implant stability was evaluated by resonance frequency analysis (Osstell AB, Gothenburg, Sweden). Data were analyzed using a two-way ANOVA followed by Tukey post hoc comparisons (α = 0.05). To facilitate interpretation of the results, empirical, non-validated clinical relevance thresholds for each outcome were defined.

Results

A total of 72 implants were placed in standardized resin models containing D3-type bone blocks (n = 12 per group). Both technology and operator experience significantly affected implant placement accuracy (p < 0.001). Compared with r-CAIS, d-CAIS resulted in higher coronal and apical deviations, with clinically relevant mean differences of 0.567 mm (95% CI: 0.396-0.738 mm) and 0.481 mm (95% CI: 0.310-0.652 mm), respectively. Clinically relevant effects of operator experience were observed only within the d-CAIS workflow, where G1 showed greater coronal (MD = 0.764 mm, 95% CI: 0.444-1.085 mm), apical (MD = 0.639 mm, 95% CI: 0.262-1.016 mm), and angular (MD = 1.502°, 95% CI: 0.737-2.267°) deviations than G3. Corresponding differences within the r-CAIS workflow did not reach predefined thresholds for clinical relevance. All groups achieved clinically acceptable ISQ values, with no clinically relevant differences attributable to the technology used or operator experience.

Conclusions

Under standardized in vitro conditions, r-CAIS achieved higher implant placement accuracy than d-CAIS. Although operator experience influenced implant placement accuracy in both workflows, differences were observed mainly within the d-CAIS workflow, while no differences were observed in primary implant stability. CLINICAL IMPLICATIONS r-CAIS achieved greater implant placement accuracy than d-CAIS and demonstrated a reduced influence of operator experience on the measured experimental outcomes.

Read PDF

Similar papers

Sep 2026

Influence of clinician experience with a static computer-assisted implant surgery workflow on implant-placement accuracy across three surgical-guide designs: a clinical trial.

OBJECTIVES Although static computer-assisted implant surgery (sCAIS) is widely used, the influence of operator experience with guided workflows-distinct from general implant experience-remains unclear. This trial compared placement accuracy between experienced and novice operators in the sCAIS workflow across three sur...

Abdelrahman K. Eldabe, Kirollos H. Botros, George N. Farid Beshay et al. · 0 citations
Open access Nov 2026

Robotic surgery in implantology: a comparative accuracy evaluation of the YakeBot system and Dcarer dynamic navigation

Objective. To compare the accuracy of single dental implant placement using the autonomous robotic system YakeBot and the Dcarer dynamic navigation system. Materials and methods. A prospective randomized clinical study of single dental implant placement was conducted from August 2025 to February 2026. The analysis inc...

S. N. Filatov, A. Muraev, K. Salekh et al. · 0 citations
Open access Sep 2026

Impact of Level of Guidance on the Accuracy of Static Computer-Assisted Implant Surgery (s-CAIS): An In Vitro Study.

Accuracy varied by guidance level and greater plan-to-implant accuracy was not accompanied by improved inter-implant parallelism, indicating that these outcomes should be evaluated separately.

Eric C. MacFadden, Chris M. Sobczak, A. Guentsch · 0 citations
Open access Nov 2026

Primary stability of dental implants placed by novice surgeons: an ex vivo study across two implant systems in a porcine mandible model

Dental implant placement is a technically demanding procedure in which surgical experience is considered a key determinant of outcome. Primary stability, as measured by the implant stability quotient (ISQ) via resonance frequency analysis (RFA), is a critical predictor of successful osseointegration. This study aimed t...

Juliane Wagner, Lais Ghafoor, Julia Baas et al. · 0 citations
Oct 2026

Surgical Efficiency, Learning Dynamics, and Positional Accuracy of Freehand Immediate Implant Placement Under Standardized Digital Planning: An In Vitro Study.

PURPOSE To characterize surgical efficiency and within-operator learning dynamics during freehand immediate post-extraction implant placement performed under standardized, prosthetically driven digital planning, and to describe the positional accuracy achieved across operators of differing experience in a 3D-printed ph...

C. Lorenzi, A. Pinto, A. Alrmali et al. · 0 citations
Sep 2026

OPTIMISING ACCURACY OF COMPUTER-ASSISTED IMPLANT SURGERY USING NOVEL DRILLING KEY DESIGN: SIMULATION-BASED STUDY.

BACKGROUND Accurate implant positioning remains the goal of guided surgery for improving surgical outcomes and minimizing complications. The aim is to compare the accuracy of standard versus novel drilling key in static-dynamic computer assisted implant surgery across D2 and D3 bone densities relative to a zero-deviati...

Thaysan Kaing, S. Arunjaroensuk, X. Yeo et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.