Analytical Performance of Vitamin D Immunoassays Compared with Electrochemiluminescence Immunoassay: A Cross-sectional Study
Abstract
Introduction: Vitamin D, the fat-soluble steroid hormone, is important in maintaining homeostasis of calcium and phosphorus. The diagnosis of Vitamin D deficiency has become considerably more frequent due to its importance in skeletal and non skeletal disorders. Comparison among immunoassay methods is essential as methodological differences can lead to variations in 25-Hydroxyvitamin D (25-OH Vitamin D) results, potentially affecting the clinical interpretation. Aim: To evaluate the analytical agreement and assay trueness of Chemiluminescence Immunoassay(CLIA) (Alkaline Phosphatase- and Acridinium Ester-based) and Latex-Enhanced Immunoturbidimetric Technique (LEIT) methods for 25-OH Vitamin D quantification using Roche e411 Electrochemiluminescence Immunoassay (ECLIA) as the reference platform. Materials and Methods: This cross-sectional study was conducted at the Department of Research and Development, Agappe Diagnostics Limited, Cochin, Kerala, India, from April 2025 to June 2025. A total of 153 randomly selected participants were included in the study. 25-OH Vitamin D quantification was performed using CLIA cartridge analyser Microprocessor-based Immunoassay Platform Analyser-i60 (MISPA-i60), CLIA cassette analyser MISPA-i180, LEIT cartridge analyser MISPA-i3, and Roche Cobas e411 ECLIA analyser as the reference platform. The strength of linear association between methods was evaluated using the Pearson’s correlation coefficient (r). Agreement between methods, systematic trend or heteroscedasticity across the measurement range were assessed using Bland– Altman analysis. Results: The cartridge-based CLEIA 25-OH Vitamin D reagent in MISPA-i60 had the best correlation with Roche e411, having a Pearson’s correlation coefficient of r=0.9612. The other two platforms, cassette-based 25-OH Vit D CLIA reagent in MISPAi180 and LEIT reagent in MISPA-i3, had good correlation regressions of r=0.9409 and r=0.9246, respectively, with the Roche e411 results. Chemiluminescence immunoassay-based platforms are more or less equal in performance, while the LEIT deviates more among the methods. Conclusion: The study observed acceptable limits of deviation among the methods of CLIA and LEIT. While all evaluated methods show strong correlation, systematic bias limits their interchangeability. It is always better to stick to the same method or even the same reagent-analyser combination for the follow-up studies of 25-OH Vitamin D quantification.