B-265 A Large Multisite Study of Beckman Coulter clinical chemistry performance demonstrates a significant majority of assays are eligible to use less QC

Oct 2026 · Clinical Chemistry · Vol 72 · 0 citations

TL;DR

The study reveals that laboratories experience a range of performance when using Beckman Coulter biochemistry instruments, implying significant practical savings can be realized through analytical Sigma metrics and subsequent QC optimization.

Abstract

The scientific literature contains sparse data on analytical method performance, especially across multiple sites There is little publicly available information about broad analytical performance of methods or instruments. To get a better understanding of assay performance, we collected routine long-term performance data and assessed it using analytical Sigma metrics From mid-2023 and through 2025, we collected and analyzed individual instrument performance for Beckman Coulter clinical chemistry assays in the US, as well as a handful in Canada. 164 instruments were assessed from 66 different laboratories. Multiple types of controls were used, from Bio-Rad to Technopath to Thermo-Fisher MAS. Peer comparison study reports from all three control vendors were accessed for each instrument. CLIA 2025 proficiency testing criteria goals were used as performance goals. Where CLIA does not set criteria, we used goals from other global sources. Cumulative mean and SD from control data were used to estimate imprecision. Peer comparison reports were used to assess bias against the cumulative peer group mean. The standard Westgard analytical Sigma-metric was calculated from the data for all available control levels. Medians and averages of the Sigma metrics were calculated. Westgard Sigma Rules were used to identify appropriate QC procedures for tests. As the analytical Sigma-metric increases, the number of rules required decreases. The study reveals that laboratories experience a range of performance when using Beckman Coulter biochemistry instruments. The median laboratory of the study population has nearly 70% of its clinical chemistry menu at 4, 5, and 6 Sigma, and thus are able to optimize and reduce QC rules. The average laboratory is almost 75% optimizable. The greatest number of methods for median and average customers is 6 Sigma. The vast majority of clinical chemistry methods in these laboratories are achieving good, excellent and world class quality. Over 91% of the assays run by the median lab are acceptable or better, demonstrating market-leading performance. Labs in the study have the opportunity to reduce their current QC practices and implement reduced Westgard Sigma Rules. The typical laboratory employs the 1:2s control rule with 2 controls and generates 9% false rejection. For 6 Sigma methods, Westgard Sigma Rules recommend switching to 1:3s with 2 controls instead, which generates less than 1% false rejection. This would reduce outliers by more than 90%, implying significant practical savings can be realized through analytical Sigma metrics and subsequent QC optimization.

View source

Similar papers

Review Oct 2026

B-266 A multisite review across the USA of major competitive biochemistry platforms reveals significant differences in analytical performance

Broad surveys of diagnostic biochemistry instruments are extremely rare. Proficiency testing surveys provide aggregated performance data, but group standard deviations are not fully reflective of individual instrument performance. The most helpful information for a clinical laboratory is to know how other individ...

S. Westgard · 0 citations
Oct 2026

Standardized Estimation of Measurement Uncertainty in Clinical Chemistry Using ISO 20914.

BACKGROUND The degree to which one is certain of results for a particular measurement/testing process is expressed as a confidence level within a range. Measurement uncertainty (MU) is a critical parameter in clinical chemistry, reflecting the degree of confidence in laboratory test results and directly influencing dia...

M. I. Ahmad, F. Ahmed, Deepa Rao et al. · 0 citations
Open access Oct 2026

A-240 Multicenter Performance Evaluation and Method Comparison of the Yumizen H550 Hematology Analyzer

Accurate hematological analysis is critical for clinical decision-making across diverse patient populations. This study evaluates the performance of the Yumizen H550, a compact hematology analyzer, by comparing it to the established Yumizen H2500 predicate device. The evaluation aimed to validate the H550’s accur...

Amin Mohammad, A. Khatibi, Joanna Hipolito et al. · 0 citations
Open access Oct 2026

Performance Assessment of Centrifugation-Based Hematocrit Testing in Primary Care Units: Evidence from Seventy-Two Proficiency Testing Datasets over Eight Years

Background: Centrifugation-based hematocrit (Hct) testing remains widely used in decentralized primary care settings where access to automated hematology analyzers may be limited. Appropriate proficiency testing (PT) performance assessment therefore requires consideration of the analytical variability associated with t...

Watcharinthon Theansun, K. Kongros, N. Apiratmateekul et al. · 0 citations
Oct 2026

A-307 Determining risk-based internal quality control frequency for biochemical testing using Sigma metrics and the QC Constellation tool

Risk-based internal quality control (IQC) has been increasingly adopted in clinical laboratories to enhance analytical quality and patient safety. ISO 15189:2022 recommends designing IQC plans based on risk assessment rather than applying fixed QC frequencies. Sigma metrics combined with the QC Constellation tool...

S. Nguyễn · 0 citations
Oct 2026

A-404 Determinants of Analytical Performance in Centrifugation–Based Hematocrit: Effects of Calibration and Internal Quality Control Frequency Across Hospital and Primary Point-of-Care Sites

Manual centrifugation–based hematocrit (Hct) remains widely used as point-of-care testing (POCT) in hospital laboratories and subdistrict health promotion hospitals (SDHHs), particularly in decentralized settings. However, evidence is limited on whether quality-related factors influence satisfactory performance w...

W. Treebuphachatsakul · 0 citations