A Projector-Based Augmented-Reality Localization Device for Breast-Conserving Surgery: Phantom Pipeline and End-to-End Error Budget

Sep 2026 · Bioengineering · Vol 13 · 0 citations · 39 references
Medicine

TL;DR

A projector-based augmented-reality (AR) localization system that projects a crosshair directly onto the patient’s skin at the predicted tumor location, which motivates phase-shifting structured light.

Abstract

Breast-conserving surgery (BCS) requires accurate intraoperative localization of the tumor, which moves between the prone diagnostic and supine surgical positions. We present a projector-based augmented-reality (AR) localization system that projects a crosshair directly onto the patient’s skin at the predicted tumor location. The pipeline combines rigid registration of ceramic fiducials between prone and supine CT (tumor error 0.30 mm on a real phantom), a SOFA finite-element simulation of gravity-driven prone-to-supine soft-tissue deformation (mean bead displacement 24.7 mm on a synthetic mesh; not validated on a deformable phantom), two-stage ICP surface alignment (RMSE 0.92 mm on a simulated camera; not yet validated on real D415 acquisitions), and projector–camera (ProCam) extrinsic calibration driving the crosshair projection. Because the tumor and sub-surface fiducials are not visible in the live RGB-D scene, a direct non-circular in vivo target registration error (TRE) on the tumor cannot be measured in the present phantom configuration. We instead report a component-based end-to-end error budget obtained as the root-sum-square of three independently measured component errors: 2.7 mm (RSS planning estimate, not a directly measured end-to-end accuracy) —rigid CT 0.30 mm, ProCam physical projection 2.50 mm, ICP 0.92 mm. The ProCam term is largest but sub-3 mm. The Gray-code reprojection residual (~11 mm in 15-pose recalibration; 8.62 mm in the deployed 4-pose calibration) overestimates the actual projection error by ~4× (resp. ~3.4×): least-squares calibration averages the consumer-DLP vertical quantization (16-level Gray-code decode on XPR-DLP) into an accurate rigid transform, so the per-observation residual reflects observation quantization, not projection accuracy. The quantization’s measurable cost is projection variance (max 5.5 mm), which motivates phase-shifting structured light. Boundary conditions on phantom rigidity, sub-surface fiducials, and simulated-camera ICP are stated explicitly.

Read PDF

Similar papers

Open access Oct 2026

Accuracy of Video Optical See-Through Augmented Reality Surgical System (VOSTARS) to Guide Craniotomy: A Phantom and a Human Cadaver Study

This study supports the technical feasibility of a patient-specific AR workflow for craniotomy guidance across anatomically different cranial approaches by enabling the translation of patient-specific preoperative planning directly to the operative field and providing AR guidance during the initial stages of cranial su...

N. Montemurro, M. Carbone, R. Viglialoro et al. · 0 citations
Open access Sep 2026

Feasibility of quantifying intra-fractional patient positional error for head and neck IMRT using radiomic features of EPID-measured fluence maps: a phantom study

This study aimed to evaluate the feasibility of estimating patient positional errors during head-and-neck fixed-field intensity-modulated radiation therapy (IMRT) using a machine-learning (ML) regression models based on radiomics features of patient-transmitted fluence maps acquired by an electronic portal imaging devi...

Takayuki Nishikata, S. Utsunomiya, Madoka Sakai et al. · 0 citations
Aug 2026

Interimplant distance and angulation measurement from CAD-generated multiunit abutment interfaces: A scan body-independent technique.

A scan body-independent technique for measuring interimplant distances and angulations directly from computer-aided design-generated multiunit abutment bottom components and captures implant spatial data complementary to existing accuracy-evaluation approaches is described.

M. Omar, Toshiki Nagai · 0 citations
Preprint Sep 2026

Object-Centered Reconstruction for Vision-Based 3D Force Estimation

A vision-based pipeline for estimating 3D interaction forces from soft-tissue deformation in stereo endoscopic video is presented and the feasibility of vision-based force estimation on an in vivo colorectal surgical sequence is demonstrated, as a step toward clinical translation of vision-based, sensorless force estim...

Zhong-Hao Zhang, Ming-Yeung Wu, Hao Yang et al. · 0 citations
Open access Sep 2026

Design, Fabrication, and Evaluation of a Patient-Derived 3D-Printed Anthropomorphic Breast Phantom for Performance Evaluation of Dual-Energy Subtraction in Contrast-Enhanced Mammography

An anthropomorphic patient-derived 3D-printed breast phantom (4.8 cm compressed thickness, 12.2% volumetric breast density) was evaluated for image-quality assessment in contrast-enhanced mammography (CEM). It comprises four 1.2 cm plates printed via fused filament fabrication (PLA/ABS). Five replicates of one of the p...

A. Belarra, Irene Hernández-Girón, P. Homolka et al. · 0 citations
Conference Open access 2026

Development Of A Mixed Reality–Based Organ Registration Assistance System For CT-Guided Puncture

Extended Reality (XR) technologies have been increasingly applied in the medical field, particularly for supporting image-guided interventions. In CT-guided biopsy, accurate puncture requires repeated CT imaging to confirm the insertion path, which results in increased radiation exposure for both patients and operators...

T. Kotone, N. Motoki, T. Ikuto et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.