Accuracy in Step Counting Is Not Accuracy in Quantifying Walking Activity: A Criterion Validity Study of Wearable Devices Under Free-Living Conditions.
Aug 2026· Medicine & Science in Sports & Exercise· 0 citations
Medicine
TL;DR
Overall, activPAL, StepWatch, ActiGraph hip/1s and ActiGraph wrist-StepCount showed the most consistent performance, which can help researchers select step-based methods that more accurately capture real-world walking patterns.
Abstract
Purpose
To determine the criterion validity of research- and consumer-grade wearables (18 step-based methods) for quantifying step counts and step-derived walking metrics in free-living conditions.
Methods
Sixty healthy adults (20-80 years, balanced by sex and decade) wore six monitors for seven days: StepWatch (ankle), activPAL (thigh), ActiGraph (hip and wrist), Fitbit Inspire 2 (hip), Fitbit Charge 5 (wrist), and Garmin Vívoactive 4S (wrist), while wearing a waist-mounted video camera at their discretion during free-living sessions. Step counts from all devices were used to derive metrics of bout duration, frequency, and intensity. Raw ActiGraph data were also processed using five wrist-based machine-learning models for step detection. Video-identified steps labeled by trained researchers served as the criterion. Equivalence testing was performed using a ±10% equivalence margin.
Results
Fifty-six participants provided 150 hours of valid video and 10,194 walking bouts. Only StepWatch fell within the equivalence margin for step counts with a mean absolute percent error (MAPE) <10%, with seven additional methods at 10-20%. For total walking bout duration, only activPAL and ActiGraph wrist-StepCount met equivalence criteria; activPAL achieved MAPE <10% and five others at 10-20%. No method met equivalence for the number of walking bouts, with MAPE >20% and balanced accuracy in bout detection <50%. For time spent in moderate-to-vigorous physical activity, activPAL (18.6%) and StepWatch (16.2%) achieved MAPE <20%, but no methods met equivalence.
Conclusions
The level of step-count accuracy obtained from wearable devices did not translate into a comparable level of accuracy in quantifying real-world walking activity. Overall, activPAL, StepWatch, ActiGraph hip/1s and ActiGraph wrist-StepCount showed the most consistent performance. These findings can help researchers select step-based methods that more accurately capture real-world walking patterns.
Accuracy improved substantially over time, although variability across brands, activities, and populations persisted; Acceptability was generally high across brands, with long wearing time and good participant satisfaction.
T. Hogue, Alexia Khoury, Pressila Njeim et al.· Frontiers in Digital Health· 0 citations
The study findings highlight the feasibility of integrating wearable HAR systems into smart homes for real-time monitoring and adaptive assistance and show superior cross-dataset generalization in Random Forest and Multilayer Perceptron models.
O. Gorjani, René Jaros, P. Bilik· Frontiers in Bioengineering...· 0 citations
The results suggest that Xiaomi Mi Band 5 provides accurate step count measurements under free living conditions, offering reliable information on overall physical activity in older adults, however, it is not sufficiently accurate for assessing calorie expenditure.
S. N. Costa, Isabela V. Scheidt, Anderson Z. Ulbrich et al.· Sport Sciences for Health· 0 citations
Wearable inertial sensors have become tools for objectively assessing mobility in older people. This study analyzed whether wearable inertial sensors could detect exercise-induced mobility changes and compared the effects of multicomponent training (MCT) and elastic band training (EBT) on mobility and physical function...
Mauricio Barramuño-Medina, Pablo Valdés-Badilla, Pablo Aravena-Sagardía et al.· Biosensors· 0 citations
The results support anthropometry-aware wearable-signal modeling as a feasible engineering approach for pediatric activity assessment, while naturalistic validation and stable calibration protocols remain necessary before routine school-based deployment.
Zhi-Yuan Yang, Liang-Wu Li, Yu-Jun Cai et al.· Medical Engineering and Phys...· 0 citations
Accelerometer-derived measures from wrist, thigh, and hip locations are strongly correlated and provide evidence to support pooled meta-analyses across wear locations and the development of location-agnostic processing software.
C. Zisou, B. Maylor, A. Acquah et al.· International Journal of Beh...· 0 citations