A-226 Benchtop Sysmex XN-450: Evaluation for Use in Primary Healthcare Setting
Abstract
Full Blood count (FBC) testing is a frequent requested investigation in primary healthcare, supporting timely clinical decision-making for a wide spectrum of patient conditions. Singhealth Polyclinics require compact analysers that can deliver reliable 5-part differential results within limited laboratory space. The Sysmex XN-450, a benchtop haematology analyser, offers a smaller footprint while maintaining advanced analytical capabilities. This study evaluates the XN-450’s performance with comparison to central laboratory’s XN-10 in a bigger floor-standing XN-3000 model. Analytical performance of the Sysmex XN-450 was evaluated in accordance with CLSI guidelines and manufacture recommendations. Studies included imprecision, linearity, limits of blank (LOB), limits of detection (LOD), carryover, accuracy, and sample stability. Routine patient samples (n=50) were analysed on the same day using both the XN-10 and XN-450 for method comparison. Data analyses consisted of regression and correlation statistics and bias plotting to compare analytical performance between the two analysers. The Sysmex XN-450 demonstrated imprecision within manufacture limits for WBC, RBC, HGB and PLT; where available, results met RCPAQAP Haematology Analytical Performance Specifications (WBC: 0.9- 2.2%, RBC: 0.5-0.9%, HGB: 0.4-0.6%, PLT: 1.6-5.7%). Linearity and carryover assessments were within manufacture-claimed limits. LOB was 0.00 for WBC, HGB and PLT, while limits LOD, carryover were 0.01 × 10^9 (WBC), 0.00 × 10^12 /L (RBC), 0.00g/dl (HGB) and 1 × 10^9 /L (PLT). Sample stability remained acceptable up to 72 hours. Accuracy was confirmed using RCPA proficiency testing materials, with all results falling within acceptable performance ranges. Method comparison with the XN-10 showed strong agreement. Passing-Bablok regression produced slopes of 0.99 to 1.10 and intercepts from - 0.52 to 0.10. Bland-Altman bias analysis demonstrated good concordance for WBC, HGB, PLT and all 5-part differentials (neutrophils, lymphocytes, monocytes, eosinophils and basophils). Operationally, the XN-450’s compact 450 × 460 mm footprint allowed easy placement on standard 700 mm depth laboratory benches, supporting its suitability for deployment across primary healthcare laboratories. In this study, Sysmex XN-450 demonstrated reliable analytical performance, meeting manufacturer specifications and external quality benchmark. Its strong agreement with the XN-10, acceptable imprecision and detection limits, and stable results over 24 hours support its suitability for routine full blood count testing in primary care. With its compact footprint and operational feasibility, the XN-450 is an appropriate replacement option for deployment across Singhealth primary healthcare laboratories.