Analytical performance of a single epitope meta-sandwich total 25-hydroxyvitamin D immunoassay and establishment of reference intervals in a Chinese population.
Abstract
Background
25-hydroxyvitamin D [25-(OH)D] concentration is considered to evaluate vitamin D status. Persistent discrepancies between chemiluminescent immunoassays (CLIAs) and liquid chromatography-tandem mass spectrometry (LC-MS/MS) continue to challenge clinical laboratories. To investigate the single epitope meta-sandwich (SEMS) immunoassay from Mindray, we performed an assessment of VD-T Ⅱ assay on the analytical performance, and the reference intervals establishment.
Methods
We performed detection limits, precision, linearity, cross reactivity studies according to American Clinical and Laboratory Standards Institute (CLSI) protocols, measured 118 selected serum samples with Mindray VD-T Ⅱ (CLIA). The results were compared with those obtained with LC-MS/MS and three other automated 25-(OH)D assays (Roche, Abbott, Snibe). Additionally, season-stratified 95% reference intervals in adults and age-specific 95% reference intervals in children of VD-T Ⅱ assay from 1402 healthy individuals are determined.
Results
The LoB, LoD and LoQ all met the claimed range; within-run and within-laboratory precision coefficient of variation (CV%) for VD-T Ⅱ was between 1.64% and 2.36%, 3.08% and 6.32% respectively; a good linearity was found in linear range from 3.00ng/mL to 150.00 ng/mL. The correlation coefficient of VD-T Ⅱ compared with LC-MS/MS was 0.963, which was the highest correlation among the assays tested; In adults, the 95% reference intervals were 7.86-48.56 ng/mL in summer, 6.28-43.34 ng/mL in winter; in children, the 95% reference intervals were 3.76-43.34 ng/mL, 6.56-50.47 ng/mL and 4.66-29.22 ng/mL in < 1 years old, 1-13 years old and 14-18 years old groups.
Conclusions
The Mindray VD-T Ⅱ assay is suitable for routine total 25-(OH)D testing.