Real-time bipolar impedance CSG-HIU overcomes historical directional limitations, achieving 100% lesion efficiency and more predictable transmurality.
Abstract
Background
High-intensity ultrasound (HIU) ablation can produce deep myocardial lesions but has been limited by the directional, side-facing transducer orientation, yielding inconsistent myocardial targeting with lesion efficiencies <75% and transmurality <60%.
Objective
We designed HIU catheters and a real-time contact sensing method to guide energy delivery, aiming for a device capable of deep ventricular tachycardia (VT) ablation and septal reduction therapy in hypertrophic cardiomyopathy (HCM). We hypothesized that contact sensing guided HIU (CSG-HIU) would create larger, deeper, and more transmural lesions than standard HIU (Std-HIU) in a terminal swine model.
Methods
Transducer dimensions (5x6 mm) and frequency (5 MHz) were selected from 3D acoustic simulations. Three contact sensing modalities (pulse-echo, and bipolar or unipolar impedance via transducer-side microelectrodes) were validated ex-vivo. The best method was then used in a prospective in-vivo swine study comparing Std-HIU (n=10) to CSG-HIU (n=13) ablation of the ventricular septum.
Results
Ex-vivo, bipolar impedance performed best, with a 600 Ω threshold showing 100% sensitivity and specificity for stable contact. In-vivo, versus Std-HIU, CSG-HIU improved lesion efficiency (100 vs 50%, p=0.007), with larger volumes (2071±1231 vs. 568±890 mm3, p=0.005), deeper lesions (15±3 vs. 6±6 mm, p<0.001), and transmurality in 69 vs 20% of swine (p=0.030). No sustained atrioventricular block, ventricular septal defects, or procedural mortality occurred.
Conclusions
Real-time bipolar impedance CSG-HIU overcomes historical directional limitations, achieving 100% lesion efficiency and more predictable transmurality.
Background:
High-intensity ultrasound (HIU) catheter ablation produces deeper lesions than radiofrequency and may serve as a new modality for interventricular septal (IVS) reduction in hypertrophic cardiomyopathy (HCM).
Hypothesis:
Transvenous HIU catheter ablation of the basal IVS will result in significant s...
Craig C. Morris, Naveed Arang, Phoebe Dijour et al.· Circulation· 0 citations
Abstract Pulsed field ablation (PFA) has rapidly emerged as a promising ablation modality for the treatment of cardiac arrhythmias because of its myocardial tissue preferential ablation and favourable safety profile. However, procedural control of lesion formation and reliable assessment of lesion durability remain imp...
M. Cespón-Fernández, R. Davalos, T. Kueffer et al.· Europace· 0 citations
Catheter ablation for scar-related ventricular tachycardia (VT) remains challenging due to limitations in ventricular mapping and the need for deeper lesions to eliminate intramural conduction channels.
To evaluate the feasibility, safety, and acute efficacy of a structured multimodal planning strategy com...
J. Cravo, D. Cazeiro, M. Vilela et al.· European Heart Journal· 0 citations
Lesion formation during pulsed field ablation (PFA) depends on catheter–tissue distance, yet no clinical parameter directly quantifies this relationship. A three-dimensional electroanatomical mapping system provides a contact index (CI) derived from impedance changes reflecting tissue proximity. The optimal CI th...
T. Ikee, T. Kamakura, Y. Ikeda et al.· European Heart Journal· 0 citations
Objective. To evaluate the thermal safety and lesion conformity of low-duty-cycle pulsed ultrasound in an ex vivo liver model adjacent to large vessels. Impact Statement. This study evaluates a low-duty-cycle pulsed ultrasound strategy that may reduce dependence on thermal dose accumulation while preserving visible ves...
Xiao Li, Ke Wu, Zhen-Xin Shen et al.· Biomedical engineering and p...· 0 citations