Phantomless calibration of glenoid trabecular bone density: improved accuracy using age-stratified reference tissue densities

Jul 2026 · JBMR Plus · Vol 10 · 0 citations · 43 references
Medicine

TL;DR

The proposed phantomless calibration approach provides a robust and clinically feasible method to evaluate glenoid trabecular bone density in both research and clinical applications and highlights the importance of considering patient-specific variations in shoulder muscle density to ensure accurate internal calibration.

Abstract

Abstract Evaluation of glenoid trabecular bone quality is important for patients undergoing shoulder arthroplasty, as the density of the glenoid bone is likely to influence implant stability. Routine CT scans show promise for quantitative bone density assessments, but CT-based calibration typically involves scanning a phantom, which limits its application in clinical practice and in retrospective research studies. Here, we developed and validated a phantomless BMD measurement technique to evaluate glenoid trabecular bone density (mg/cc) using clinical shoulder CT scans of 88 patients (48 males, 40 females; mean age: 53 years). Air, fat, and muscle tissue were used as internal density references to develop patient-specific calibration functions for glenoid BMD estimation. The phantomless calibration was validated by comparison with an asynchronous phantom-based calibration using the European Spine Phantom. A Bland-Altman analysis yielded a mean difference of 0.37 mg/cc (95% CI: [–1.39, 2.14] mg/cc) between the two methods. Further improvement in internal calibration accuracy was achieved by implementing an age-stratified approach (two age groups; cut-off 55 years) to account for age-related declines in muscle density. With age stratification of reference tissue densities, the mean difference between internal and phantom-based calibration was reduced to 0.20 mg/cc (95% CI: [–1.38, 1.78] mg/cc). The proposed phantomless calibration approach provides a robust and clinically feasible method to evaluate glenoid trabecular bone density in both research and clinical applications. Moreover, the results highlight the importance of considering patient-specific variations in shoulder muscle density to ensure accurate internal calibration.

Read PDF

Similar papers

Open access Oct 2026

Methods for Computed Tomography‐Based Glenoid Bone Loss Measurement Show High Interobserver Consistency but Variable Accuracy

Purpose To evaluate the reliability of computed tomography (CT)‐based glenoid bone loss calculation methods in shoulder instability by assessing interobserver agreement using 3‐dimensional virtual bone models and to identify the technique that provides the most optimal balance for clinical application. Methods Between...

Ayyappan V. Nair, Aravind Rajan, Kiran Veerendra et al. · 0 citations
Open access Sep 2026

Normative Glenoid Bone Mineral Density Values in a Non-Pathologic Adult Population: A Quantitative Computed Tomography Study with Age and Sex Stratification

These locally derived normative HU values, obtainable opportunistically from routine thoracic CT without additional radiation, may assist radiologists and shoulder surgeons in preoperative bone-quality assessment, implant selection, and baseplate/screw trajectory planning for anatomic and reverse total shoulder arthrop...

Mamta Verma · 0 citations
Open access Oct 2026

Reliability of Implant‐ vs. Bone‐Based Postoperative Measurements After Anatomic Total Shoulder Arthroplasty Using CT Imaging

In anatomic total shoulder arthroplasty (aTSA) research, postoperative anatomical measurements are often referenced to the reamed glenoid bone surface because the glenoid component is not fully visible on CT. However, the reliability of measurements based directly on the implant (implant‐based method) and their agreeme...

N. Daneshvarhashjin, P. Debeer, F. Verhaegen et al. · 0 citations
Open access Nov 2026

CT-derived Hounsfield units of the acromion and scapular spine: Age- and sex-related associations and measurement reliability in adults.

BACKGROUND Acromial and scapular spine fractures are important complications after reverse total shoulder arthroplasty (rTSA). Although CT-derived Hounsfield units (HU) permit site-specific assessment of regional bone quality, reproducible measurement sites remain uncertain. We examined associations of acromial and sca...

Yuki Kurisu, Kiminori Yukata, Yuji Saeki et al. · 0 citations
Open access Oct 2026

Acute glenoid rim fractures: reliability of linear and surface area methods for CT-based bone loss assessment

We wanted to investigate and compare the reliability of two measuring methods for glenoid bone loss, one linear- and one surface area method, in acute glenoid fractures after dislocation. This was an experimental study. Thirty CT-shoulder scans in patients with an acute glenoid rim fracture were measured in a lateral”e...

Katarina Lindgren, L. Adolfsson, A. Stålman et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.