A scintillator-based approach to spatial and temporal dosimetry in UHDR electron irradiation
Abstract
The clinical translation of FLASH-radiotherapy remains limited by the lack of accurate dosimetry at ultra-high dose rates (UHDR), as many conventional detectors exhibit strong dose-rate dependence or insufficient temporal resolution. Scintillation detectors, with their fast response and high spatial capability, represent a promising alternative. In this study, a scintillator-camera system designed to provide pulse-resolved temporal and high-resolution spatial dosimetry for pulsed UHDR electron beams was evaluated. A thin-film scintillator was irradiated using a modified 10 MeV clinical linear accelerator delivering an average dose rate of ≈200 Gy/s at 200 Hz pulse repetition frequency. The system demonstrated excellent linearity (R2 = 0.998) across the investigated DPP range (0.06–2.1 Gy). The duration of the scintillator afterglow (∼1 ms) was well below the 5 ms interpulse interval, ensuring clean temporal separation of pulses. Pulse-by-pulse spatial profiles showed high reproducibility, enabling reliable assessment of beam symmetry and flatness. These findings demonstrate that scintillator-based imaging can provide accurate, pulse-resolved UHDR dosimetry with simultaneous 2D spatial information.