A dual-phase study of a bioabsorbable venous vascular closure system: identifying predictors of failure and ensuring long-term safety in over 1,000 procedures

Oct 2026 · European Heart Journal · Vol 47 · 0 citations

TL;DR

A dual-phase evaluation confirms that the bioabsorbable VVCS is highly effective, although its efficacy is influenced by BMI and sheath size.

Abstract

As the volume of catheter ablation for arrhythmias increases, optimizing post-procedural hemostasis is essential for clinical workflow and patient comfort. The VASCADE MVP is an extravascular, bioabsorbable collagen-based venous vascular closure system (VVCS). We conducted a dual-phase study to identify predictors of procedural failure and evaluate long-term safety, specifically focusing on the management of rare infectious complications. This study, conducted at our hospital, utilized a two-phase design. Phase 1 involved a mechanistic analysis of 203 consecutive venous access sites (70 patients) to identify predictors of hemostatic failure. Phase 2 consisted of an expanded safety evaluation of 1,162 consecutive procedures through October 2025. For atrial fibrillation ablation, an uninterrupted DOAC protocol was used with ACT targets of >300s (RF/cryoballoon) or >350s (pulsed-field ablation). Post-deployment, institutional protocol required 3–4 hours of gauze compression before ambulation. In Phase 1, the procedural success rate in the initial cohort was 94%. Multivariate analysis identified female sex (P=0.04), lower BMI (22.7 ± 3.5 vs. 25.0 ± 3.6 kg/m², P=0.02), and large-bore sheaths (10.4 ± 1.8 vs. 8.9 ± 1.2 Fr, P=0.01) as significant predictors of failure. In Phase 2, the safety cohort (n=1,162), delayed access-site cellulitis occurred in 0.6% (7/1,162) approximately one-month post-procedure. Management of these infections was limited to oral antibiotics and minor local management (incision, irrigation, and local debridement). Crucially, no patient required major surgical intervention, such as vascular reconstruction or bypass grafting, which are sometimes reported with suture-mediated closure devices. This dual-phase evaluation confirms that the bioabsorbable VVCS is highly effective, although its efficacy is influenced by BMI and sheath size. The system’s primary safety advantage is its bioabsorbable nature, which allows rare infectious complications to be managed conservatively, thereby avoiding the invasive surgical interventions associated with permanent suture-based devices.Table

Read PDF

Similar papers

Open access Oct 2026

One thousand cases of Pulsed-Field Ablation: Real-World insights

Pulmonary vein isolation (PVI) is the standard catheter-based treatment for atrial fibrillation (AF). Due to its safety profile and efficacy, pulsed field ablation (PFA), a non-thermal energy, has been increasingly adopted as the default strategy across many centers. We aimed to analyse the real-world efficacy an...

S. Azevedo, R. Carvalho, J. Pereira et al. · 0 citations
Open access Sep 2026

ARC-VAC Study (Analysis of real-world multicenter experience with the large-bore CLEANER-Vac Thrombectomy System): An analysis of 200 patients

Background Venous thromboembolism (VTE), encompassing deep vein thrombosis and pulmonary embolism, represents a significant clinical entity associated with elevated morbidity and mortality rates. Conventional therapeutic approaches, such as systemic anticoagulation and thrombolytic agents, frequently yield suboptimal t...

Sameh Sayfo, C. Ishak, Ilan Rzadkowolsky-Raoli et al. · 0 citations
Open access Oct 2026

Long-term efficacy of pulsed-field ablation devices in real-world atrial fibrillation patients: a direct comparison between the Farapulse and the Affera ablation systems

Pulsed-field ablation (PFA) with its characteristic high tissue-selectivity, has recently emerged as a promising alternative to the conventional radiofrequency ablation for the management of atrial fibrillation (AF). This study evaluated the safety and efficacy of two systems using either a pentaspline or a latti...

S. Mohanty, V. L. La Fazia, C. Gianni et al. · 0 citations
Open access Oct 2026

Comparison of the Technical Success of Obtura and StarClose Systems for Femoral Arterial Access-Site Closure in Neurointerventional Procedures: A Single-Center Retrospective Cohort Study

Background/Objectives: To compare technical hemostasis outcomes of collagen-based Obtura and nitinol clip-based StarClose vascular closure devices (VCDs) for femoral access-site closure after transfemoral neurointerventional procedures in two cohorts treated during non-overlapping periods. Methods: In this single-cente...

N. Kat, Tirdad Setayeshi · 0 citations
Open access Oct 2026

Safety and effectiveness of a mid-voltage pulsed-field ablation catheter in persistent atrial fibrillation: acute outcomes

High-voltage pulse field ablation (PFA) can disrupt red blood cell membranes, leading to an increased risk of hemolysis. A novel small-balloon, mid-voltage PFA system is expected to reduce the risk of hemolysis through lower pulse voltages while achieving favorable acute success rates. To evaluate the acut...

B. Ren, R. Zeng, X. Hu et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.