B-355 A High-Throughput LC-MS/MS Method Maintains Equivalent Analytical Performance in Evaluation of Liver Function by the HepQuant DuO® Test
Abstract
Chronic liver disease (CLD) is estimated to affect more than 1.5 billion people worldwide and account for approximately 2% of worldwide deaths. The HepQuant DuO® test is a minimally invasive Laboratory Developed Test (LDT) that measures liver function by assessing the clearance of isotopically labelled cholate in patient serum, providing a Disease Severity Index (DSI) to support diagnosis and clinical management. Herein, we describe a high-throughput LC-MS/MS method for evaluation of liver function. This study aims to develop and evaluate the feasibility and performance of this high-throughput LC-MS/MS method using native patient samples with diverse disease etiologies, and to compare the results with those obtained using the original, longer DuO LDT method. The original DuO LDT uses a 6-minute gradient, a longer Agilent Zorbax Eclipse C8 column (4.6 x 150 mm, 5 µm), and a 1.0 mL/min flow rate. The high-throughput method reduces the gradient to 3 minutes, employs a shorter Agilent Poroshell 120 EC C8 column (3.0 x 50 mm, 2.7 µm), and a 0.8 mL/min flow rate. The analytical performance characteristics including accuracy, precision, linearity, matrix effect, and interference were assessed between the two methods. Five model patient samples, with varying etiologies from Hepatitis C Virus (HCV), Non-Alcohol related Steatohepatitis (NASH), and alcohol related liver disease (ALC) were extracted in triplicate by three analysts, over three days, and analyzed by two instruments (Sciex API 4500 LC-MS/MS). Inter and intra-day variability were assessed for the new method using ANOVA statistical analysis. Model patient DSIs were compared between the high throughput, 3-minute method, and the original, 6-minute method using Bland-Altman and paired t-test analyses. The high-throughput methodology reduced the runtime from 6 minutes to 3 minutes. The accuracy of the assay ranged from 90%-106% between the two methods. No matrix effects were observed for either method. Inter- and intra-day variability was constant between analysts, days, and instruments used with no statistically significant difference. DSIs ranged from 10.6 to 39.3, where DSI values above 18.3 indicate an increased risk of Large Esophageal Varices (LEVs). The high-throughput method yielded DSI values comparable to the original 6-minute method, with no significant statistical or clinical differences; DSI differences ranged from -1.73% to 1.07%. Increasing throughput for the DuO LDT enables reduced analytical turnaround time to quantitative liver function data. The high-throughput method performs equivalently to the original, with no significant differences in accuracy, matrix effects, interference, variability, or DSI values. Application of the new method approximately doubles analytical throughput in a typical LC-MS/MS run and improves turnaround times from accessioning of samples to reporting results.