Point-of-care detection of C-peptide in diabetes management: a novel electrochemical immunosensor based on magnetic iron-oxide nanoflowers.
Abstract
A novel electrochemical immunosensor for point-of-care detection of C-peptide based on a graphite screen-printed electrode (GP-SPE) modified with magnetic iron-oxide nanoflowers (IONFs) has been developed for diabetes management. The surface structure, morphology and electrochemical properties of unmodified and modified electrodes were investigated by scanning electron microscopy (SEM), transmission electron microscopy (TEM), energy-dispersive X-ray spectroscopy analysis (EDX), X-ray photoelectron spectroscopy (XPS), Raman spectroscopy, cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS), demonstrating the enhanced electrochemical response of the platform modified with magnetic IONFs compared to iron-oxide nanoparticles (IONPs) and to unmodified electrode. The resulting IONFs/GP-SPE platform exhibited competitive analytical performances, including a wide dynamic range (0.25-4 nM), a low detection limit (70 pM) and excellent selectivity. Lastly, the biosensor has been successfully tested in human serum and urine samples, demonstrating an optimum agreement with a conventional HPLC-UV method.