Validation of the AViTA upper arm blood pressure monitor BPM636 in a combined pediatric and adult population according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2018/Amd2:2024.

Oct 2026 · Blood Pressure Monitoring · 0 citations · 13 references
Medicine

TL;DR

The AViTA BPM636 passed the criteria of the ANSI/AAMI/ISO Universal Standard protocol and can be recommended for home-based blood pressure measurements in the tested population, including children aged 5-12 years and adults aged 21-82 years, with arm circumferences ranging from 15 to 42 cm.

Abstract

Objective

To evaluate the clinical accuracy of the AViTA BPM636 oscillometric upper-arm blood pressure monitor in the tested pediatric and adult population, including children aged 5-12 years and adults aged 21-82 years with arm circumferences ranging from 15 to 42 cm, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) Universal Standard (ISO 81060-2:2018/Amd2:2024).

Methods

The AViTA BPM636 was evaluated in 85 participants (including 35 pediatric and 50 adult subjects) with upper-arm circumferences ranging from 15 to 42 cm, and two cuffs of different sizes (15 to 22 and 22 to 42 cm) were used in this study. The recruitment fulfilled the requirements of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) for a combined pediatric and adult population using the same arm sequential blood pressure measurement method.

Results

For criterion 1, the mean ± SD of the differences between the test device and reference blood pressure measurements was 0.93 ± 6.11 mmHg for SBP and -0.20 ± 5.64 mmHg for DBP. For criterion 2, the SD of individual mean differences for SBP (5.44 mmHg) and DBP (4.98 mmHg) remained well within the maximum permissible limits of 6.88 and 6.95 mmHg, respectively, thereby satisfying the validation criteria.

Conclusion

The AViTA BPM636 passed the criteria of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) protocol. It can be recommended for home-based blood pressure measurements in the tested population, including children aged 5-12 years and adults aged 21-82 years, with arm circumferences ranging from 15 to 42 cm.

Read PDF

Similar papers

Open access Oct 2026

Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.

The ARM-30F-XL complies completely with International Standard ISO 81060-2:2018+Amd 2:2024 and is recommended for BP measurement in individuals with an arm circumference of 22-52 cm for home and clinical use.

Yong-Hua Qiu, Wei Han, Li-Jun Liu et al. · 0 citations
Sep 2026

Clinical validation of the MED-58 upper-arm automated oscillometric blood pressure monitor with fan-shaped and rigid cuffs for home and clinical use in adolescents and adults, according to International Organization for Standardization 81060-2:2018 (Amd.1:2020; Amd.2:2024): a single-center prospecti

This study validated the accuracy and safety of the MED-58 upper-arm automated oscillometric blood pressure monitor - intended for home and clinical use - with fan-shaped and rigid cuffs, per International Organization for Standardization (ISO) 81060-2:2018 (Amd.1:2020; Amd.2:2024). In this prospective, single-center v...

Hong-Yun Ma, Jing-Zhen Liu, Meng-Ya Chen et al. · 0 citations
Open access Oct 2026

Accuracy Evaluation of the Omron HEM‐7120 Automated Oscillometric Upper‐arm Home Blood Pressure Monitor in Children and Adolescents Using the AAMI/ESH/ISO 81060–2:2018 Numerical Criteria

This study evaluated the Omron HEM‐7120 automated oscillometric upper‐arm home blood pressure monitor in children and adolescents against the numerical accuracy criteria of the AAMI/ESH/ISO Universal Standard. Test readings were compared with mercury auscultatory readings from two physicians using the same‐arm sequenti...

Tipanan Pisithkul, Kwanchai Pirojsakul · 0 citations
Open access Oct 2026

Comparison of the 2024 ESC and 2025 AHA/ACC classifications of elevated blood pressure with regards to prevalence and blood-pressure associated cardiac remodelling in the general population

The 2024 ESC and 2025 AHA/ACC guidelines on elevated blood pressure and hypertension offer different classification systems. There is limited evidence directly comparing the association of these classifications with echocardiographic changes in the general population. We aimed to understand the differences...

A. Ohlrogge, E. Unger, L. Hannen et al. · 0 citations
Sep 2026

Clinical validation of MESI mTABLET BP according to ISO 81060-2:2018 + A1 + A2.

OBJECTIVE The objective of this study was to clinically validate the MESI mTABLET BP device in accordance with the protocol outlined in ISO 81060-2:2018 + A1 + A2. METHODS A total of 127 adult participants were enrolled, of whom 36 were excluded. The final study sample consisted of 91 participants with upper arm circ...

Jaka Strel, Metka Križič, Biljana Nikolova Stoilova et al. · 0 citations
Open access Oct 2026

Performance of a 15-Day Continuous Glucose Monitoring System in Children and Adolescents With Diabetes

Background: The CareSens Air (CSAir) continuous glucose monitoring system is approved for use in adults in South Korea and received adults-only CE marking in early 2024. The aim of this prospective, multicenter, pivotal study in children and adolescents with insulin-dependent diabetes mellitus was to determine if the p...

Manuela Link, C. Denzer, T. Biester et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.