CPRE64-CZT: A Low-Noise 64-Channel Front-End ASIC With Energy and Timing Readout for CdZnTe Detectors
Abstract
This article presents CPRE64-CZT, a low-noise 64-channel front-end readout application-specific integrated circuit (ASIC) for cadmium zinc telluride (CdZnTe or CZT) detector systems that require configurable-polarity cathode and anode readout. Each channel integrates a low-noise charge-sensitive amplifier (CSA), a high-order shaper, a peak-detect-and-hold (PDH) circuit for energy measurement, and a sample-and-hold circuit for timing measurement, with power dissipation below 5 mW per channel. The three-side pad arrangement supports compact detector coupling and enables expansion to 128 channels with limited area overhead. With a biased CZT detector connected, the measured equivalent noise charge (ENC) is below 150 e<inline-formula> <tex-math notation="LaTeX">${}^{-}$ </tex-math></inline-formula> rms for the anode channels and below 250 e<inline-formula> <tex-math notation="LaTeX">${}^{-}$ </tex-math></inline-formula> rms for the cathode channel. The readout chain achieves an integral nonlinearity (INL) within 1.5% over a maximum input charge range of 350 fC. In <sup>137</sup>Cs measurements with a pixelated CZT detector, depth-dependent energy correction using the cathode-to-anode (C/A) ratio improved the summed 662-keV energy resolution to 1.00% FWHM, with the best-performing pixel reaching 0.78% FWHM. These results demonstrate low-noise spectroscopic readout with energy and timing observables suitable for subsequent detector-level depth correction and position-sensitive CZT analysis.