B-011 Evaluation of the Tosoh HLC-723GR01 HbA1c Analyzer Using Normal Hemoglobin Samples and in the Presence of Common Hemoglobin and Thalassemic Variants

Oct 2026 · Clinical Chemistry · 0 citations

Abstract

HbA1c is used to aid in the diagnosis of diabetes, identify individuals at risk for diabetes, and monitor long-term glycemic control. Venous whole blood collected in K2EDTA tubes from normal hemoglobin (HbAA) specimens (n=135) spanning 4.2%–16.6% HbA1c was tested as both whole blood and corresponding hemolysates (n=135). HbAA specimens were analyzed on the HLC-723G8 in triplicate and on the HLC-723GR01 as single measurements at three user-representative sites. Interference from common hemoglobin variants was assessed using HbAS (n=29), HbAC (n=25), HbAD (n=22), and HbAE (n=34), and from thalassemic variants using elevated HbA2 (n=33) and HbF (n=33), by comparing results to the HLC-723G8. For HbAA specimens, relative bias across HbA1c decision points ranged from 0.0% to 0.6% for both whole blood and hemolysates. Total error (bias plus imprecision) was =2.21% for whole blood and =2.49% for hemolysates across 5%–12% HbA1c. For variant specimens, relative bias ranged from -2.6% to 2.8%, meeting the acceptance criterion of =±7% at clinically relevant HbA1c levels (6.5% and 8.0%). Chromatographic separation supported consistent identification of the evaluated hemoglobin variants. The HLC-723GR01 demonstrated accurate HbA1c results for HbAA specimens and for samples containing common hemoglobin and thalassemic variants, with biases within predefined acceptance limits.