Nov 2026· Hormone Research in Paediatrics· Vol 99· 0 citations
TL;DR
There was a negative overall bias for F and D sensors (higher GMI compared to HbA1c) leading to a relative increase in insulin delivery in an AID pump.
Abstract
HbA1c has been the gold standard for glycemic outcomes. Increasing emphasis is now placed on continuous glucose monitoring (CGM). CGM data over several weeks can be converted to the Glucose Management Indicator (GMI), which estimates the HbA1c expected if glucose control remains stable over 2–3 months.
To investigate the relationship between HbA1c and sensor metrics with different CGM systems.
Data from the most recent visit in the Swedish National Diabetes Register (extracted up to 2023, < age 18) were analyzed. Mean CGM glucose over the preceding 14 days was used to calculate GMI. Analyses were performed separately for 3 sensor systems: Dexcom G6 (D), FreeStyle Libre 3 (F), and Guardian 4 (G).
5,789 individuals (85.7% used pumps, 14.3% multiple injections) with type 1 diabetes were included. Mean HbA1c was 50.4 mmol/mol (6.8%), 50.1 for D, 51.6 F, 51.5 G. Mean glucose was 8.5 mmol/l (154 mg/dl), GMI 52.9 mmol/mol (7.0%). HbA1c 48 mmol/mol (6.5%) represented a mean glucose of 8.4 (D), 8.4 (F) and 7.9 mmol/l (G), and TITR (Time in Tight Range, 3.9-7.8 mmol/l, 70-140 mg/dl) of 50.3 (D), 51.3 (F) and 57.6% (G). Fig 1 shows Bland-Altman plots.
In automated insulin delivery (AID) systems, insulin dosing is guided by sensor glucose values. Therefore, systematic differences between sensors may lead to variations in HbA1c despite similar mean glucose levels. In this study, there was a negative overall bias for F and D sensors (higher GMI compared to HbA1c) leading to a relative increase in insulin delivery in an AID pump. The bias was rather similar over the span of HbA1c values for F, while D and to an even greater extent G sensors had a negative bias at lower HbA1c levels and a positive at higher. This would result in lower mean glucose at higher HbA1c levels, leading to a relative decrease in AID insulin delivery. These biases should be considered when interpreting CGM metrics and advising patients on AID insulin pump use.
In most participants, HbA1c exceeded GMI, revealing a predominantly unidirectional discrepancy whose magnitude surpassed previously reported rates, consistent with the proportional bias observed and the high baseline glycemic burden of this cohort, with direct implications for therapeutic decision-making.
S. Lemos, G. L. Torres, L. P. S. Lima et al.· Hormone Research in Paediatr...· 0 citations
AIM
The glucose management indicator (GMI) estimates HbA1c from continuous glucose monitoring (CGM)-derived mean glucose. We assessed whether this relationship differs between contemporary CGM systems.
MATERIALS AND METHODS
In this observational study, mean sensor glucose was derived from 2 months of CGM data precedi...
Katrine Bovién Gørlitz, Sara Heebøll, M. Fenger-Grøn et al.· Diabetes, obesity and metabo...· 0 citations
AIMS
The continuous glucose monitoring (CGM)-derived glucose management indicator (GMI) is used for estimating glycated haemoglobin (HbA1c). This study assessed agreement between GMI-calculated using both original and updated (uGMI) models-and laboratory-measured HbA1c across Dexcom, MiniMed and Abbott sensors.
MATER...
Lavanta Farouk, Bilal Bashir, R. Seese et al.· Diabetes, obesity and metabo...· 0 citations
Objective. To evaluate the effectiveness of continuous glucose monitoring (CGM) compared to point-of-care (POC) glucose testing in reducing HbA1c among adults with diabetes. Methods. This single-center, retrospective cohort study included adult patients with type 1 or type 2 diabetes attending Bayhealth Sussex Endocrin...
Eshita Sharmin, Cameron Golden, Ejaz Mahmood· Delaware journal of public h...· 0 citations
Therapeutic patient education (TPE) is central to pediatric type 1 diabetes (T1D) management. Its impact on dynamic glycemic metrics — time in range, variability, and circadian patterns — remains underexplored. This study evaluates the combined effect of structured TPE and continuous glucose monitoring (CGM) with...
D. Senouci, A. S. Bendeddouche· Hormone Research in Paediatr...· 0 citations
Continuous glucose monitoring systems (CGM) has become an integral component of type 1 diabetes mellitus (T1DM) management by improving glycemic monitoring and reducing diabetes burden. Since January 2025, CGM systems have been reimbursed by the Turkish Social Security Institution, enabling wider use among childr...
K. Çetin, Y. Bahar, A. Tırtar et al.· Hormone Research in Paediatr...· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.
See our Cookie Policy.