Early and midterm outcomes of the Gore Excluder Iliac Branch Endoprosthesis within and outside the instructions for use for aortoiliac and iliac artery aneurysms.
Abstract
Objective
To evaluate perioperative and midterm outcomes of endovascular aneurysm repair using the Gore Excluder Iliac Branch Endoprosthesis (IBE) for aortoiliac and iliac artery aneurysms treated within and outside the instructions for use (IFU).
Methods
This single-center retrospective cohort study included patients who underwent IBE implantation for non-ruptured aortoiliac or iliac artery aneurysms between April 2017 and December 2022. Outcomes were analyzed at the patient, procedure, or IBE-treated-side level, as appropriate. The primary endpoints were technical success, IBE-related complications, and reinterventions. Secondary endpoints included internal iliac artery (IIA) branch patency, aneurysm sac status, and freedom from reintervention or aneurysm-related death.
Results
A total of 56 unique patients underwent 58 procedures involving 73 IBE-treated sides; 24 sides were within-IFU and 49 were outside-IFU. Technical success was achieved in 100% of within-IFU procedures and 95.0% of outside-IFU procedures. At 36 months, freedom from reintervention was 79.1% and 94.0% in the within-IFU and outside-IFU groups, respectively (P = .24), and IIA branch patency was 88.8% and 85.5%, respectively. No significant between-group difference in IIA branch patency was detected after accounting for within-patient clustering (hazard ratio, 1.53; 95% confidence interval, 0.28-8.35; P = .62). Seven IIA branch occlusions occurred during follow-up, none of which was associated with buttock claudication. In exploratory univariable analysis, type II endoleak was associated with CIA aneurysm sac growth (odds ratio, 7.86; 95% confidence interval, 1.08-57.0; P = .04).
Conclusions
IBE use outside the IFU was feasible in carefully selected patients. No significant differences in midterm clinical outcomes were detected between the within-IFU and outside-IFU groups.