A-136 Improving HbA1c Result Reportability Using Dual-Method Verification: Bio-Rad D-100 HPLC and Siemens Healthineers DCA Immunoassay
Abstract
Hemoglobin A1c (HbA1c) is a critical biomarker for diabetes diagnosis and monitoring long-term glycemic control. Separation-based methods, such as the Bio-Rad D-100 ion-exchange HPLC and the Sebia Capillarys® Octa 3 capillary electrophoresis system, provide excellent resolution of hemoglobin species and variants, which is essential for accurate HbA1c quantification. However, chromatographic limitations, including peak overlap, peak tailing, and the appearance of unknown peaks that contribute to area calculations, may prevent accurate assignment or integration of the HbA1c peak. These issues trigger instrument flags and lead to non-reportable results despite the clinical need for reliable HbA1c values. When separation is compromised, employing a method with a fundamentally different analytical principle may improve accuracy. Immunoassays, such as the Siemens Healthineers DCA Vantage, target the glycated N-terminal valine of the ß-globin chain directly and do not rely on chromatographic resolution, providing an orthogonal approach to verify results generated by separation-based systems. This study assessed a dual-method verification workflow using the DCA Vantage Analyzer and independently evaluated the Sebia Capillarys® Octa 3 to enhance reportability of Bio-Rad D-100–flagged samples. Flagged HbA1c results from the Bio-Rad D-100 were retested using the Siemens Healthineers DCA Vantage latex-enhanced turbidimetric inhibition immunoassay. Agreement between methods was assessed by calculating the relative percent difference; results were considered verified if the DCA and D-100 values agreed within ±8%, consistent with CLIA total allowable error criteria for HbA1c. Verified results were reported without modification. Samples exceeding ±8% difference remained invalid and were accompanied by interpretive comments indicating potential hemoglobin variant interference and recommending fructosamine testing as an alternative glycemic marker. Separately, a subset of Bio-Rad–flagged samples was analyzed using the Sebia Capillarys® Octa 3. This capillary electrophoresis platform separates hemoglobin fractions based on electrophoretic mobility and may produce separation patterns distinct from ion-exchange HPLC. The primary outcome for both analyses was the proportion of previously flagged samples that produced reportable, unflagged HbA1c results using either orthogonal testing approach. A total of 85 Bio-Rad-flagged samples underwent verification with the Siemens DCA method. Of these, 47 samples (55%) met the ±8% agreement criterion and were verified for reporting. The remaining 38 samples (45%) demonstrated larger discrepancies and were considered as invalid. In the Sebia Capillarys® Octa 3 evaluation (n = 69), 51 samples (74%) yielded reportable, unflagged HbA1c results. The remaining 18 samples (26%) remained non-reportable. Compared with immunoassay verification, the Sebia method produced a higher percentage of reportable results and offered direct visualization of hemoglobin variants. Separation-based HbA1c methods provide strong analytical performance but may be occasionally limited by chromatographic artifacts that obscure the HbA1c peak. Incorporating an orthogonal immunoassay method enhances reliability, because it measures the glycated site directly. Combining immunoassay verification with a separation-based method significantly improves reportability and provides greater analytical assurance when evaluating samples with potential hemoglobin variant interference or chromatographic artifacts.