A-137 Incidence of Abnormal Hemoglobin Expression in a Large HbA1c Sample set on CAPILLARYS 3 and Potential Impact on Clinical Reporting of HbA1c
Abstract
HbA1c% remains the primary diabetes screening and long-term monitoring analyte as it mirrors the long-term damage done by protein glycation under chronically elevated glucose levels, as demonstrated by the landmark DCCT study. However, this testing assumes a patient with normal hemoglobin expression; predominantly Hb A with an average red blood cell lifespan of 120 days. Patients with hemoglobinopathies challenge this expectation: hemoglobin variants and thalassemias may interfere with HbA1c% measurement and/or disrupt the formation or accumulation of HbA1c. In extreme cases, patients may express no Hb A, in which case ADA standards dictate that the Hb A1c cannot and should not be used in patient evaluation. Knowing when patients have relevant hemoglobinopathies is essential to proper HbA1c evaluation. The Sebia CAPILLARYS 3 HbA1c assay produces electropherograms illustrating patient hemoglobin expression and incidentally detects many variants and thalassemias. A complete five month database of HbA1c testing on CAPILLARYS 3 instruments at Advocate Clinical Laboratories - Atrium Core Laboratory (Charlotte, North Carolina) was extracted and de-identified for evaluation. The data set comprised 166,883 total runs, with 162,187 unique samples. Results were categorized according to flags and values on the CAPILLARYS 3 system, indicating abnormal hemoglobin profiles. Abnormal results were reviewed for presumptive identification of hemoglobin variants or thalassemic states. Out of the 162187 samples run, 135712 (83.7%) showed no abnormal flags, 25,451 (15.7%) were flagged as potentially having a hemoglobin variant, 19,129 (11.8%) were indicative of some form of thalassemia or iron deficiency anemia (increased/decreased Hb A2, etc.), including 1627 (1%) with only abnormally high Hb A2. 157 (0.1%) were found to have profiles characteristic of patients who should not have been tested with an Hb A1c assay (suspicion of transfusion, absence of Hb A, S/B+ thalassemia, etc.) Of the 26,475 abnormal electropherograms, 1,514 had only clinically insignificant amounts of Hb F or delta globin variants; reportable without further comment. A further 23,979 samples could likely be reported with a cautionary note that the patient expressed a significant variant or thalassemia that could potentially impact their accumulation of HbA1c, and the result should be closely correlated with the patient’s chart. The remaining results represented more uncertain cases which might benefit from a hemoglobinopathy panel for variant identification and insight into whether it might influence HbA1c interpretation. In the subject data set, while the majority of samples would be regarded as having “normal” hemoglobin expression, or containing minor abnormalities with little or no clinical impact on HbA1c, a significant minority of cases were found in which the presence of a variant or thalassemia may grant greater insight into the patient’s clinical condition and should be taken into consideration when their A1c is evaluated. A much smaller proportion of results were found to be unsuitable for HbA1c testing due to the absence of native Hb A, or indicative of conditions known to result in rapid RBC turnover and would require alternate testing to be considered.