Skip to content

CLOSING THE LOOP – APPLYING REFERENCE MEASUREMENT SERVICES TO EXTERNAL QUALITY ASSURANCE FOR ROUTINE CLINICAL CHEMISTRY TESTS

2025 · 150th anniversary of the Metre Convention — From Units to the Universe · 0 citations

TL;DR

Correctly performed RMS testing of commutable EQA materials adds supporting information that manufacturers are supplying traceable assays, and supplies evidence for methods that may need improvement, and can support identification of poorly performing methods.

Abstract

Introduction Low between-method bias is vital for interpretation of medical laboratory tests for patient care. For common clinical chemistry tests the manufacturers are responsible for providing reagents, instruments and calibrators which can provide traceable results. The Royal College of Pathologists of Australasia Quality Assurance Program (RCPAQAP) has run an external quality assurance program with presumed commutable materials – The Liquid Serum Chemistry Program - for over 10 years. Currently over 270 laboratories are enrolled measuring up to 60 measurands with 2 or 4 samples per year. In 2022 and 2023 two samples each year were sent for value assignment in JCTLM-listed Reference Measurement Services (RMS). We assess the added utility of performing this RMS testing. Methods Freshly collected serum samples are frozen and rapidly distributed. Laboratories submit results to the RCPAQAP together with supporting information including instrument, reagent manufacturer and method principle. Samples were also sent to either RFB in Germany or WEQAS in Wales for RMS testing for 14 measurands. The following instrument groups were assessed: Abbott Alinity and Architect, Beckman-Coulter AU, QuidelOrtho Vitros, Roche Cobas, and Siemens Atellica and Dimension. Bias assessment was made against the overall method median and against RMS values, using criteria based on Australian common reference intervals, biological variation or expert recommendations. Results The effect of performing the RMS testing added to result interpretation in the following ways. For some tests with acceptable between-method bias for all methods, it was shown that all methods were also unbiased to the RMS results (ALT(+P5P), ALP, creatinine, urate). For tests where one or more method was biased relative to the all-assay median, that most methods had acceptable bias and the outlier method(s) were biased against the RMS result (AST(+P5P) – Dimension low; CK – Vitros low; Cholesterol – Beckman high; Amylase – Dimension low; Calcium – Dimension low; Magnesium – Atellica Low; Triglycerides – Vitros, Roche High). For GGT and Glucose all methods showed a bias to the RMS (negative and positive respectively). For ALP, LDH and cholesterol there was a change in the bias for all methods relative to the RMS between 2022 and 2023. For ALP all methods remained within acceptable bias, but for LDH, while between-method bias was acceptable, for 2022 data several methods exceeded bias limits against the RMS. These inconsistent biases, or all-method biases raise concerns about analyte stability and sample handling for RMS testing which is generally delayed and further afield. Discussion Reference method value assignment for presumed commutable EQA samples can increase confidence that methods working correctly, and can support identification of poorly performing methods. The Measurement Uncertainty of RMS results can be relevant in assessing method bias. Limitations to providing RMS testing include sample stability; required sample volumes; the limited number samples included; and costs. We note that in this study we included only 14 of the measurands in the program. Conclusions Correctly performed RMS testing of commutable EQA materials adds supporting information that manufacturers are supplying traceable assays, and supplies evidence for methods that may need improvement.

View source

Similar papers

Review 2025

The European Metrology Network TraceLabMed : coordinating activities of European NMIs/DIs and calibration laboratories in the field of laboratory Medicine

A coherent strategy for a joint European response to the need for establishing and verifying results metrological traceability in accordance with the IVD Regulation (EU) 2017/746 is created.

V. Delatour, G. O'Connor, Rainer Stosch et al. · 0 citations
Open access Aug 2026

An Update on State-of-the-Art Performance for Precision in Medical Laboratories.

BACKGROUND The European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) has established a three-stage model of criteria to set analytical performance specifications (APS). The third model is based on state-of-the-art (SoA), which has been described as the highest level of analytical performance with cur...

Jérémie Gras, Delphine Collin-Chavagnac, Florian Scherrer et al. · 0 citations
Open access Sep 2026

Verification of Patient Result Comparability Among Nine Hematology Analyzers Using the CLSI EP31-A-IR Guideline: A Four-Year Experience.

INTRODUCTION Clinical laboratories operating multiple analyzers with random sample allocation must ensure result comparability across instruments to support consistent clinical interpretation and patient safety. OBJECTIVES To evaluate the applicability of the CLSI EP31-A-IR guideline for periodic verification of pati...

Xavier Tejedor-Ganduxé, Fernando Marqués-García, Cristina Martínez-Bravo et al. · 0 citations
Open access Aug 2026

Model 1b analytical performance specifications – report on the Prague 2025 workshop

Abstract The main purpose of establishing APS is to ensure that test performance aligns with their specific clinical goals. Analytical performance specifications serve multiple functions: as acceptance criteria to verify that in vitro diagnostic medical devices (IVD MD) meet regulatory and quality standards before mark...

A. Horvath, Sverre Sandberg, M. Thelen et al. · 0 citations
Open access Aug 2026

Validation of point-of-care INR testing against standard laboratory methods in a Malaysian primary healthcare setting

Reliable INR monitoring is crucial for patients on warfarin to prevent thrombotic or bleeding events. This study evaluated the accuracy of the Wondfo POCT device compared to standard laboratory methods (Stago STA-R Max3 and ACL TOP 300) in a primary healthcare setting. The goal was to assess whether this point-of-care...

Rosline Hassan, S. Iberahim, S. S. Mohd Yusoff et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.