Abstract 13783: Hemodynamic Comparison of a Novel Prosthetic Valve and the Native Mitral Valve in a Pulsatile Flow Model
Abstract
Background: Current prosthetic mitral valves, including mechanical and bioprosthetic valves, often provide a smaller effective orifice area than the native mitral valve, potentially impairing left ventricular filling and limiting cardiac output after mitral valve replacement. We developed a novel cylinder-type prosthetic valve designed to preserve physiologic transmitral flow. This study evaluated whether the novel valve reproduces native mitral valve hemodynamics across a wide range of heart rates. Methods: Porcine native mitral valves (n=12) and novel prosthetic valves (n=12) were evaluated using a pulsatile mock circulatory loop. Hemodynamic performance was assessed at heart rates of 70, 100, and 130 bpm while maintaining a target mean flow of 3 L/min. Systolic blood pressure (SBP), diastolic blood pressure (DBP), forward flow, and net flow were measured and compared between valve types using two-way ANOVA. Results: The novel prosthetic valve demonstrated a bicuspid closing configuration resembling that of the native mitral valve. Across all tested heart rates, hemodynamic parameters were comparable between the novel prosthetic and native mitral valves. There was no difference between valve types in SBP/DBP at 70, 100, and 130 bpm with 119±7.1/91±5.7, 106±4.8/90±3.7, and 97±5.1/83±4.5 mmHg for native valves, compared with 119±7.9/91±2.9, 106±5.5/89±4.1, and 99±7.9/85±6.6 mmHg for novel prosthetic valves (all p>0.05). SBP significantly differed by heart rate but no significant effect valve type-heart rate interaction was observed for SBP or DBP. Forward flow was also comparable between native and prosthetic valves at all heart rates (70 bpm: 3.7±0.4 vs. 3.6±0.2 L/min; 100 bpm: 3.5±0.5 vs. 3.4±0.2 L/min; 130 bpm: 3.3±0.6 vs. 3.3±0.6 L/min), with no significant valve-type effect or interaction (p=0.324 and p=0.954).Similarly, net flow was preserved with the prosthetic valve and was comparable to native valve performance at 70 bpm (3.1±0.4 vs. 3.0±0.2 L/min), 100 bpm (2.8±0.4 vs. 2.7±0.2 L/min), and 130 bpm (2.4±0.5 vs. 2.4±0.4 L/min), without significant differences between valve types or interaction effects (p=0.210 and p=0.978). Conclusions: The novel cylinder-type prosthetic valve reproduced native-like hemodynamics across physiologic and tachycardic heart rates in the ex vivo study. Preservation of blood pressure, forward flow, and net flow suggests that this design maintains transmitral flow after mitral valve replacement. These findings support the potential of this next-generation prosthetic valve to address a fundamental limitation of conventional mitral valve prostheses and provide more physiologic postoperative hemodynamics.