Oct 2026· Frontiers in Surgery· Vol 13· 0 citations· 22 references
Medicine
TL;DR
Reimplantation is a salvage procedure that successfully restores auditory function after device failure, allowing for continuous language development and Electrode selection and revision strategy should remain tailored to each case.
Abstract
Introduction Cochlear implantation (CI) is a highly reliable intervention for sensorineural hearing loss; however, device failure may occur, limiting patient benefit unless revision surgery is performed. This study evaluated the long-term institutional prevalence and causes of CI failure, and assessed complications, audiological and speech outcomes, and their associated factors following multiple revision surgeries. Materials and methods A retrospective cohort study identified 91 failed CI cases among 3,777 CIs performed at a single tertiary referral center between January 1993 and November 2024. Audiological and speech outcomes, including Pure Tone Average (PTA), Speech Reception Threshold (SRT), Speech Discrimination Score (SDS), Categories of Auditory Performance (CAP), and Speech Intelligibility Rating (SIR), were analyzed using Mixed Models for Repeated Measures (MMRM) with restricted maximum likelihood (REML) estimation. Results The overall prevalence of CI failure was 2.41%, occurring predominantly in pediatric patients (96.7%). Hard device failure was the most common cause (68.8%), followed by medical (15.0%), soft (13.8%), and surgical (2.5%) failures. Revision surgery was performed once in 80 cases, twice in 8 patients, and three times in 2 cases. Reimplantation successfully preserved audiological performance across successive revisions, while CAP and SIR scores demonstrated significant additional improvement after the first revision compared with the initial implantation (p = 0.037 and p < 0.001, respectively). Reimplantation using the same electrode model was associated with better SDS after the first revision (p = 0.035), and better SRT (p = 0.022) and SIR (p = 0.001) after the second revision. An interval of >3 years between surgeries was associated with significantly higher post-revision SIR scores after both the first (p = 0.040) and second (p = 0.004) revisions. Conclusion CI failure had a prevalence of 2.41% over more than three decades. Reimplantation is a salvage procedure that successfully restores auditory function after device failure, allowing for continuous language development. Electrode selection and revision strategy should remain tailored to each case.
BACKGROUND
Although cochlear implants (CIs) effectively restore hearing, device abandonment ('non-use') remains a critical challenge limiting long-term auditory rehabilitation.
AIMS/OBJECTIVES
This study aimed to evaluate the prevalence, clinical and radiological characteristics, and patient-reported reasons for CI n...
Jeong Hun Jang, Jungho Ha, You-Jeong Kim et al.· Acta Oto-Laryngologica· 0 citations
BACKGROUND
Cochlear implantation (CI) is standard for severe-to-profound sensorineural hearing loss, but outcomes in children with cerebral palsy (CP) remain unclear.
AIMS/OBJECTIVES
To compare auditory and speech outcomes after CI in children with and without CP.
MATERIAL AND METHODS
This retrospective cohort incl...
T. Kazemi, S. Nabavizadeh, Seyed Basir Hashemi et al.· Acta Oto-Laryngologica· 0 citations
OBJECTIVES
To evaluate outcomes following cochlear implantation (CI) in postlinguistically deaf adults with prolonged auditory deprivation.
METHODS
A retrospective analysis was conducted on 95 patients who underwent CI between 2020 and 2022. Inclusion criteria were age ≥18 years. Main outcome measures included the pu...
M. Lopes, H. Figueiredo, Francisca Bartolomeu et al.· Cochlear Implants Internatio...· 0 citations
OBJECTIVE
To compare audiological, speech, and functional auditory outcomes following simultaneous bilateral cochlear implantation (SBLI) and early sequential bilateral cochlear implantation (SQLI) in children, and to explore the relationship between inter-implant interval and outcomes in sequentially implanted childre...
Lokanath Sahoo, V. Gaurav, R. Chugh et al.· International Journal of Ped...· 0 citations
OBJECTIVES
To evaluate the safety and efficacy of cochlear implantation for auditory rehabilitation in patients with nasopharyngeal carcinoma (NPC) after radiation.
DATA SOURCES
CINAHL, Cochrane Library, PubMed, and Scopus were searched from inception through February 2026.
REVIEW METHODS
Studies reporting cochlear...
Kaitlyn A. Roberts, Lauren E. Williamson, Shaun A. Nguyen et al.· The Laryngoscope· 0 citations
OBJECTIVE
To evaluate the impact of prolonged unaided duration of deafness (DoD) on speech perception outcomes following cochlear implantation in adults.
METHODS
A retrospective review was performed of adult cochlear implant recipients at a single high-volume center between 2012 and 2023. Patients with ≥ 5 years of d...
J. Cottrell, Michael Scudder, Emily R. Spitzer et al.· The Laryngoscope· 0 citations