Calibration of X-ray multimeters for half-value layer: measurement methods and uncertainty evaluation

Oct 2026 · Metrologia · Vol 63 · 1 citation · 29 references
Physics

Abstract

The first half-value layer (HVL), defined as the thickness of material required to reduce the x-ray beam intensity (air kerma) to 50%, is a key radiation quality specifier. Although HVL has been measured for decades using ionization chambers and aluminum filters, the detailed measurement procedures and associated uncertainties remain insufficiently established. With the growing use of x-ray multimeters (XMMs) capable of measuring HVL directly from the x-ray beam, reliable calibration and uncertainty evaluation are needed. This work presents a detailed approach for determining the first HVL and calibrating XMMs for RQR 2, RQR 5, and RQR 10 standard radiation qualities for radiography as defined in IEC 61 267. The uncertainty budget is structured into three steps: measured HVL, reference HVL, and HVL calibration coefficient, emphasizing the effects of ionization chamber energy dependence and field size. The reference HVLs for RQR 2, RQR 5, and RQR 10 were 1.417 mm Al, 2.570 mm Al, and 6.510 mm Al, respectively, with expanded uncertainties (k = 2) of 1.0%, 1.1%, and 1.7%. The corresponding HVL calibration coefficients had uncertainties of 1.4%, 1.3%, and 1.8%.