Calibration errors and their impact on measurement accuracy in lifetime based phosphor thermometry
Abstract
This study investigates the error propagation in the calibration stage of lifetime based phosphor thermometry and establishes a quantitative framework linking calibration errors to final temperature uncertainty. The results demonstrate that the primary error source originates not from the lifetime measurement, but from composite calibration errors, contributing over 99% of the total uncertainty. Through derivation of error evaluation equations for linear, quadratic, and cubic calibration curves, the dominant role of the covariance coefficient and residuals in uncertainty propagation is elucidated. Experimental validation shows only 3.5% difference in total uncertainty between high and low noise conditions, confirming that environmental noise has minimal impact compared to inherent systematic bias in calibration models. This study ultimately demonstrates that substantial improvements in accuracy must follow a clearly defined optimization hierarchy: first, the measurement accuracy of the reference thermocouple must be ensured; on this basis, the residual term can then be reduced by narrowing the calibration range, selecting an appropriate calibration equation, and/or increasing the density of calibration points.