Testing Efficiency Transfer software in gamma spectrometry: a comparison of different calculation methods for increased complexity of sample geometries.
Abstract
Background
Gamma spectrometry with high-resolution High-Purity Germanium detectors is a widely adopted technique, valued for its excellent energy resolution. It is applied across a range of Radiation Protection activities, including the measurement of low-activity concentrations in environmental samples containing natural and artificial radionuclides, as well as quality assurance assessments of radiopharmaceuticals used in diagnostic imaging.
Methods
This work presents an inter-comparison between software for Efficiency Transfer calculations, including three semi-empirical (EFFTRAN, ETNA, and ANGLE) and two Monte Carlo based codes (GESPECOR and LabSOCS/ISOCS). The calculations considered sample geometries of increasing complexity, ranging from a Marinelli beaker to a vial and wipe test. A verification test (IAEA-TERC-2022-01/02) was conducted to assess the accuracy of the simulations by validating the results against a Verification Sample with well-known low activity radionuclide concentrations.
Results
Calculations performed for the different sample geometries yielded consistent Efficiency Transfer coefficients across all the software tested. The largest relative bias was observed at a low photon energy (59keV from 241Am), which exceeded the 20% acceptability limit for the more complex sample under investigation. A similar trend was observed in the activity concentration results obtained from the verification test.
Conclusions
A comprehensive evaluation of the most used Efficiency Transfer software was conducted to assess their performance in gamma-ray spectrometry. This work aims to enhance the accuracy of activity concentration estimates, for samples with non-standard geometries, thus enabling rapid and reliable responses in potential radiological emergencies.