Validation of Minimally Invasive and Noninvasive Hemodynamic Monitoring in Adolescents: The Pediatric Cardiac Output Monitoring Observational (POGO) Study.
Aug 2026· Pediatric Critical Care Medicine· Vol 27 8, pp.
986-997
· 0 citations· 36 references
Medicine
TL;DR
The minimally and noninvasive CO measurements with FloTrac and ClearSight APCO show acceptable bias and precision under standardized elective clinical conditions in an adolescent pediatric population, making them useful tools in the routine clinical management of critically ill children.
Abstract
Objectives
Minimally invasive cardiac output (CO) monitoring technologies lack validation in the adolescent population. We aimed to test the accuracy and precision of the FloTrac arterial pressure cardiac output (APCO) and ClearSight photoplethysmography cardiac output (PPGCO; Edwards Lifesciences, Irvine, CA) technologies in adolescents undergoing elective cardiac catheterization.
Cardiac catheterization laboratory at three freestanding pediatric hospitals in the United States.
PATIENTS
Adolescent patients 12-18 years of age undergoing surveillance cardiac catheterization after heart transplantation.
Interventions
Swan-Ganz pulmonary artery catheterization with the intermittent thermodilution CO (iCO) technique was used as the gold standard for CO measurements. APCO was measured using an existing femoral arterial sheath, and PPGCO was measured utilizing a volumetric cuff placed on the finger of the upper extremities. Two sets of simultaneous iCO, APCO, and PPGCO measurements were compared. Bias and level of concordance were assessed using Bland-Altman analysis.
Measurements
AND MAIN
Results
Out of the 89 subjects enrolled, 73 underwent measurements and provided a total of 60 sets of measurements for each device. The median age was 16 years (interquartile range [IQR], 14-17 yr) with a median body surface area of 1.6 m2 (IQR, 1.5-1.8 m2). The median averaged intermittent cardiac index was 2.7 L/min/m2 (IQR, 2.3-3.1 L/min/m2). FloTrac APCO mean bias was 0.32 L/min/m2 (IQR, 0.20-0.44 L/min/m2) with a precision of 0.70 L/min/m2 (IQR, 0.58-0.82 L/min/m2). ClearSight PPGCO mean bias was 0.28 L/min/m2 (95% CI, 0.16-0.41 L/min/m2) with a precision of 0.71 L/min/m2 (95% CI, 0.59-0.84 L/min/m2). No serious adverse events were reported related to the devices.
Conclusions
The minimally and noninvasive CO measurements with FloTrac and ClearSight APCO show acceptable bias and precision under standardized elective clinical conditions in an adolescent pediatric population, making them useful tools in the routine clinical management of critically ill children.
Due to the clinical importance of early detection of hemodynamic compromise and the limitations of transthoracic echocardiography (TTE) as an intermittent, operator-dependent assessment, the introduction of non-invasive cardiac output monitoring (NICOM) in the neonatal population has generated considerable interest. El...
L. van Wyk, S. Martini· Seminars in Fetal & Neonatal...· 0 citations
BACKGROUND
The Venous Excess Ultrasound (VExUS) score is increasingly used to assess systemic venous congestion, yet direct validation against invasively measured right atrial pressure (RAP) in pulmonary hypertension (PH) remains limited.
RESEARCH QUESTION
How accurate is the VExUS score to predict RAP in patients wi...
M. D'Alto, N. Cangiano, Antonio Orlando et al.· Chest· 0 citations
OBJECTIVES
To validate an ultrasound-based method for noninvasive estimation of central venous pressure (CVP) in critically ill patients and compare its performance with invasively measured CVP.
METHODS
We conducted an analytical cross-sectional study in adults admitted to a general intensive care unit. Patients with...
Carmelo José Espinosa-Almanza, Jorge Leonardo Támara-Rivera, Héctor Andrés Ruiz-Ávila· Journal of ultrasound in med...· 1 citation
BACKGROUND
Echocardiography is widely used for non-invasive measurements of stroke volume and cardiac output, while Ultrasonic Cardiac Output Monitor (USCOM) offers an alternative Doppler-based approach. Whether these two modalities can be used interchangeably in clinical practice remains unclear. This study aims to ev...
OBJECTIVES
Volume assessment remains a challenging bedside practice that is highly dependent on patient factors, comorbid conditions, physical exam, and point-of-care ultrasound (POCUS) techniques. POCUS is increasingly used to assist with diuretic decision-making for patients with suspected volume overload. While deve...
Aaron M Silver, Michael Armstrong, Jean M. G. Sabile et al.· Journal of ultrasound in med...· 2 citations
ABSTRACT Backgrounds: Extubation failure after pediatric cardiac surgery increases morbidity and intensive care unit (ICU) stay. Conventional indices such as rapid shallow breathing index (RSBI) and PaO2/FiO2 do not account for multi-organ contributors such as cardiac function, pulmonary aeration, and diaphragmatic per...
R. Gopalakrishnan, Rohini Dhar, Sandra Chaithanyam Bijukumar et al.· Annals of Pediatric Cardiolo...· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.