Directional Radial Force Alignment Predicts Pacemaker Implantation After Balloon-Expandable Transcatheter Aortic Valve Implantation: A Computed Tomography-Based Biomechanical Study.

Jul 2026 · American Journal of Cardiology · Vol 275, pp. 27-30 · 0 citations · 10 references
Medicine

TL;DR

The Vectorial Angle is a novel, independent pre-procedural predictor of PPM after balloon-expandable TAVI, highlighting that the spatial orientation of the valve's maximal radial force relative to the conduction tissue is a potential anatomical determinant associated with conduction-system vulnerability, complementing conventional anatomical assessments such as MSL.

Abstract

Background

Conduction disturbances requiring permanent pacemaker (PPM) implantation remain a frequent complication of transcatheter aortic valve implantation (TAVI). This study aimed to evaluate the "Vectorial Angle"-a novel computed tomography (CT)-derived biomechanical parameter-alongside traditional anatomical and electrocardiographic predictors of PPM following balloon-expandable TAVI

Methods

This retrospective cohort study included 67 patients with severe symptomatic aortic stenosis undergoing TAVI with the balloon-expandable Myval system. The Vectorial Angle was defined on pre-procedural CT at the elliptical left ventricular outflow tract (e-LVOT) level as the angular deviation between the minor axis (representing the axis of maximal radial expansion force) and the midpoint of the right coronary to non-coronary (RCC-NCC) commissure (approximating the anatomical course of the conduction system).

Results

PPM implantation was required in 12 patients (17.9%). In multivariate logistic regression analysis, a narrower Vectorial Angle (OR 0.935 per 1° increment, 95% CI 0.889-0.982; p=0.007) and a shorter membranous septal length (MSL) (OR 0.447 per 1 mm increment, 95% CI 0.241-0.831; p=0.011) emerged as the only independent predictors of PPM requirement. Valve oversizing index was not associated with PPM implantation (p=0.954). Implantation depth was numerically greater in the PPM group but did not reach independent predictor status (p=0.079). Pre-existing right bundle branch block (RBBB) lost independent predictive value after adjustment for these structural parameters. Receiver operating characteristic (ROC) curve analysis for the Vectorial Angle yielded an area under the curve of 0.806 (p=0.001), with an optimal cutoff of ≤52°.

Conclusions

The Vectorial Angle is a novel, independent pre-procedural predictor of PPM after balloon-expandable TAVI. This finding highlights that the spatial orientation of the valve's maximal radial force relative to the conduction tissue is a potential anatomical determinant associated with conduction-system vulnerability, complementing conventional anatomical assessments such as MSL.

View source

Similar papers

Aug 2026

Conduction Disturbances and Permanent Pacemaker Implantation After TAVR With Self-Expanding Devices: The CONNAVIPRO Study.

BACKGROUND The Evolut and the Navitor transcatheter heart valves (THVs) represent the most widely used self-expanding implanted valves. Despite a similar implant concept, they markedly differ in design and hemodynamic performance. In both THVs, conduction disturbances following transcatheter aortic valve replacement (T...

Enrico Poletti, O. Oliva, Ahmad Hassan et al. · 0 citations
Open access Feb 2026

Transcatheter aortic valve-in-valve implantation for failed surgical aortic bioprostheses using the Navitor platform: early outcomes in a consecutive single-center cohort

Introduction Transcatheter heart valve implantation into a failed surgical heart valve (THV-in-SHV) is an established reintervention strategy for degenerated aortic bioprostheses; however, early procedural outcomes, discharge hemodynamics, and standardized safety remain clinically relevant in routine practice. Material...

R. Sobczyński, J. Trębacz, Paweł Kleczyński et al. · 0 citations
Aug 2026

Newer-generation balloon-expandable vs. self-expanding transcatheter aortic valves: a propensity score matched analysis.

Among patients treated with newer-generation transcatheter valves, SEV demonstrated more favorable hemodynamic performance compared with BEV, however, short- and mid-term mortality rates remained similar between the two groups.

M. Baudo, S. Sicouri, Dimitrios E. Magouliotis et al. · 1 citation
Oct 2026

Early outcomes of the Octapro balloon-expandable transcatheter heart valve in an unrestricted cohort of patients with aortic valve stenosis: the MyPro registry.

BACKGROUND Data on early outcomes of the novel balloon-expandable (BE) Octapro transcatheter heart valve (THV) in real-world practice are limited. We sought to assess procedural safety and early outcomes of the Octapro in patients with severe native aortic stenosis (AS) undergoing transcatheter aortic valve implantatio...

A. Ielasi, E. Moscarella, Gregor Verček et al. · 0 citations
Open access Sep 2026

The Cusp-Overlap Technique Versus the Standard Three-Cusp Technique for Self-Expanding Transcatheter Aortic Valves—A Single-Center Experience

Background: Conduction disturbances and the subsequent need for permanent pacemaker implantation (PPMI) remain a major limitation of transcatheter aortic valve replacement (TAVR) with self-expanding transcatheter heart valves (SE-THVs). The cusp-overlap technique (COT) has been proposed to optimize implantation depth a...

A. Ispas, D. Brunet, R. Saady et al. · 0 citations
Open access Sep 2026

Preprocedural computed tomography angiography aortic root geometry and 1-year left ventricular reverse remodeling after transcatheter aortic valve replacement

Introduction and objectives Left ventricular reverse remodeling (LVRR) after transcatheter aortic valve replacement (TAVR) is associated with outcomes, but the relevance of aortic-root geometry remains uncertain. We examined whether the annular eccentricity index (EI) and sinus of Valsalva height on computed tomography...

Xiao-Ting Li, Dai-Song Jiang, Jian Liu et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.