Evaluation of a Wearable Device for Cardiorespiratory Monitoring in Healthy Anaesthetised Cats

Jul 2026 · Veterinary Medicine and Science · Vol 12 · 0 citations · 17 references
Medicine

TL;DR

In healthy cats, the OSS demonstrated high agreement with the reference monitor for HR measurements and provided clinically acceptable SpO2 values within established equivalence limits.

Abstract

ABSTRACT Objective The aim of this study was to evaluate the accuracy and agreement of heart rate (HR) and peripheral oxygen saturation (SpO2) measurements obtained using the wearable Owlet Smart Sock (OSS) in anaesthetised cats, compared with a reference veterinary bedside monitor (Monitor 1). Methods Twenty healthy cats aged 1–2 years and weighing 3.2 and 5.1 kg were enrolled. All animals underwent a standardised general anaesthesia protocol. HR and SpO2 were recorded simultaneously using the OSS and Monitor 1 at 1‐minute intervals for 20 min. A total of 307 paired measurements were analysed. Agreement was assessed using Bland–Altman analysis and clinical equivalence was evaluated using the two one‐sided tests (TOST) procedure, with predefined equivalence margins of ±10 bpm for HR and ±3% for SpO2. Results A very strong correlation was observed between OSS and Monitor 1 for HR measurements (r = 0.995; p < 0.0001). The mean difference was 0.18 bpm (95% CI: −0.38–0.40), demonstrating clinical equivalence (p < 0.0001). For SpO2 measurements, a statistically significant linear correlation was identified (r = 0.634; p < 0.0001). The mean difference was ‐0.5% (95% CI: −0.74 to −0.29), remaining within the predefined ±3% equivalence margin and indicating clinical equivalence (p < 0.0001). Conclusion In healthy cats, the OSS demonstrated high agreement with the reference monitor for HR measurements and provided clinically acceptable SpO2 values within established equivalence limits. Further validation under diverse clinical conditions and in different patient populations is warranted.

Read PDF

Similar papers

Aug 2026

Assessing the agreement and reliability of Magene® for heart rate monitoring.

To evaluate the accuracy and reliability of the Magene® heart rate (HR) monitor compared to the Polar® H7 (validated device) during rest, maximal graded exercise test, and active recovery. A within-group study was conducted with 68 healthy adults (21 women). Participants completed two laboratory sessions using both mon...

Mila Alves Matos Rodrigues, Thiago Danillo Silva, Jéssika Teodoro Santos et al. · 1 citation
Open access Sep 2026

Assessing respiratory rates and signs of air hunger using the PneumoRator sensor.

Objectives: To investigate whether the PneumoRator wearable sensor could accurately assess Respiratory Rate (RR) and signs of air hunger (SAH) compared with clinician-observed assessment in adults with and without breathing pattern disorder (BrPD). Design: A pilot cross-sectional observational study. Setting: A researc...

J. Hudson-Colby, H. Baston, M. Harding et al. · 0 citations
Open access Aug 2026

Accuracy, safety, and usability of a real‐time continuous glucose monitoring system in pediatric diabetes management

ABSTRACT Background The purpose of the study was to evaluate the accuracy, safety, and usability of a real‐time Continuous Glucose Monitoring (CGM) system compared with venous blood glucose (vBG) in children and adolescents. Methods The study enrolled 81 participants aged from 3 to 17 at three sites in China. Two senso...

Bing-Yan Cao, Z. Su, Yi Wang et al. · 0 citations
Open access Sep 2026

VALIDATION OF HEART RATE MEASUREMENTS FROM THE POLAR VERITY SENSE DURING THE SUBMAXIMAL EKBLOM-BAK TEST IN ADULT INDIVIDUALS WITH OBESITY

Objective To assess the validity of Polar Verity Sense, an arm-worn heart rate (HR) monitor, for measuring HR in individuals with obesity during the submaximal ergometer cycle Ekblom-Bak test. Subjects/Patients Adults in an inpatient rehabilitation programme for individuals with obesity. Methods Participants completed...

Grete Rudi Onstad, Fredrik Klæboe Lohne, M. S. Fimland · 0 citations
Open access Sep 2026

A flexible piezoelectric sensor for non-contact monitoring of respiratory and heart rates: accuracy and clinical feasibility under simulated dental treatment conditions

Vital signs are not routinely monitored in dental practice, potentially delaying the recognition of medical emergencies. This study aimed to evaluate the accuracy of a piezoelectric sensor (Taido Sensor, Sumitomo Riko Co., Ltd., Nagoya, Japan) for measuring respiratory rate (RR) and heart rate (HR). Resilience against...

F. Kozu, Maho Suyama, Hiroyuki Kurakami et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.