Jul 2026· Journal of Endovascular Therapy· pp.
15266028261467575
· 0 citations· 17 references
Medicine
TL;DR
In a patient with a symptomatic pararenal aneurysm undergoing urgent endovascular repair, electrified wire in situ fenestration (EWISF) proved challenging in a patient with a symptomatic pararenal aneurysm undergoing urgent endovascular repair.
Abstract
Purpose
Electrified wire in situ fenestration (EWISF) represents an effective off-the-shelf option in time-sensitive endovascular procedures. This article highlights the importance of radiofrequency energy delivery duration for successful catheter crossing with subsequent ballooning and bridging stent placement.
CASE REPORT
An 89-year-old man with symptomatic pararenal abdominal aortic aneurysm was transferred to our endovascular center. Given the patient's advanced age and the urgent clinical scenario, an endovascular approach was preferred. Pre-stenting of the stenotic left renal artery (LRA) was followed by intentional coverage of both renal arteries using a tubular aortic endograft with suprarenal fixation (Endurant). The right renal artery was revascularized using a bare metal stent (Visi-Pro) via the parallel technique, whereas the LRA was targeted with EWISF. Intraoperatively, catheter crossing after initial EWISF proved highly challenging. A repeat EWISF procedure with prolonged radiofrequency application was required to accomplish revascularization. Distal extension with a bifurcated endograft, appropriately sized limbs, and subsequent implantation of a physician-modified thoracic endograft, to manage insufficient overlap between the previously implanted endografts, completed the operation.
Conclusion
Electrified wire in situ fenestration is effective for endograft modification; prolonged energy delivery may be required to achieve catheter passage.Clinical ImpactThe electrified wire technique has emerged as a valuable alternative for in situ fenestration in urgent cases, enabling target vessel revascularization at low cost using readily available equipment. In a patient with a symptomatic pararenal aneurysm undergoing urgent endovascular repair, electrified wire in situ fenestration (EWISF) proved challenging. After intentional coverage of the renal artery ostium with a tubular aortic endograft, the initial EWISF attempt did not allow catheter passage through the endograft. A second attempt with prolonged electrical energy application ultimately enabled successful revascularization of the renal artery. Application duration may affect EWISF success, and acknowledgment of this parameter seems essential for enhanced technique reproducibility.
Central venous stent migration is a rare complication in hemodialysis patients. While antegrade migration toward the heart or pulmonary arteries has been documented, the mechanism and management of retrograde migration toward the distal axillary vein remain unreported. We present a 61-year-old woman with stenosis at th...
Ai-Qiang Zhou, Liang-Zhu Hu, Yong Lu et al.· Journal of Vascular Access· 0 citations
PURPOSE
To report the first application of the Leak-Loop Technique (LLT) that combines controlled endoleak and dual guidewire loop stabilization to address cerebral perfusion maintenance and endograft malpositioning during in situ fenestrated thoracic endovascular arch repair (fTEVAR) under urgent resource-limited sett...
Ye Tian, K. J. Kwan, Jin Li et al.· Journal of Endovascular Ther...· 0 citations
(1) Background: Laser in situ fenestration (L-ISF) is an endovascular technique used to preserve target vessel patency in urgent or emergency endovascular aortic repair when fenestrated/branched solutions are unavailable or unsuitable, or as a bailout technique for fenestration or branch misalignment. This on-bench stu...
T. Ormandzhieva, Florian Elger, Marcello Silvano et al.· Bioengineering· 0 citations
Wide-neck middle cerebral artery (MCA) aneurysms incorporating acutely angulated branches remain among the most technically challenging lesions in neuroendovascular therapy. Selective branch catheterization and subsequent catheter exchange may be difficult using conventional techniques.
A 79-year-old man wit...
BACKGROUND
Dedicated bridging stents for fenestrated and branched endovascular aortic repair (F/BEVAR) have recently become available. The BeFlared stent (Bentley InnoMed, Germany) is a balloon-expandable covered stent mounted on a dual-diameter balloon designed to allow simultaneous deployment and flaring. This study...
Tamas Kovacs, Andreas K. Johannessen, Jacob Budtz-Lilly et al.· Journal of Cardiovascular Su...· 0 citations
Guidewire entrapment and fracture is an uncommon but potentially hazardous event during percutaneous coronary intervention (PCI), and may lead to thrombosis, vessel occlusion, and embolization. In clinical practice interventional cardiologists are aware that such complication is not that rare, especially in complex PCI...
Pietro Mazzarotto, Delio Tedeschi, Simone Morganti et al.· Catheterization and cardiova...· 0 citations