Jul 2026· Ear and Hearing· 0 citations· 32 references
Medicine
TL;DR
Compared with unaided listening, hearing aids negatively affected externalization and own-voice perception, with larger effects for closed than open fittings, however, at the individual level, measures of occlusion and sound externalization were not strongly related.
Abstract
Objectives
A recent study found that hearing aids fitted with highly occluding domes had an impact on the perception of externally presented speech, producing sound images that were perceived as too close or even inside the head. The present study examined whether this phenomenon is related to the well-known occlusion effect, which describes changes in ear canal levels and perceived own-voice quality during speaking.
Design
Participants were 10 young adults with normal hearing and 10 young adults with sensorineural hearing loss. All participants were fitted with both closed- and open-fit hearing aids, and hearing aid users were also tested with their own devices. Externalization was measured using a numerical scale from 0 (inside the head) to 10 (at the loudspeaker). Participants also rated the naturalness of their own voice while speaking using a subjective scale from 1 (most unnatural) to 10 (most natural). Occlusion was also measured objectively for open and closed fittings using real ear measurements.
Results
Externalization ratings were significantly higher for unaided than aided listening and for open versus closed domes. Similarly, own-voice ratings tended to be higher for unaided versus aided listening and for open versus closed domes. Although these broad patterns suggest that reduced ratings are associated with more occlusion, objective measures of occlusion did not account for a significant amount of the variance in either kind of rating. Similarly, individual own-voice ratings did not significantly predict individual externalization ratings. The data obtained with hearing aid users wearing their own devices suggested that regularly worn devices can have positive effects on own-voice perception while disrupting externalization.
Conclusions
Overall, compared with unaided listening, hearing aids negatively affected externalization and own-voice perception, with larger effects for closed than open fittings. However, at the individual level, measures of occlusion and sound externalization were not strongly related.
Evidence is found that utilizing passive flat-attenuating ER-20 earplugs, and active flat-attenuating MP9-15 earplugs provide better speech understanding in noise for normal hearing listeners.
Trevor A. Simones, Darryl M. Horn, M. Pienkowski· 0 citations
OBJECTIVE
In this study, we aimed to compare open-set environmental sound perception performance and response patterns across four groups-hearing aid (HA) users, cochlear implant (CI) users, young normal-hearing adults, and older normal-hearing adults-and to examine background factors associated with environmental soun...
Atsuya Takayama, Kei Sakamoto, Han Matsuda et al.· Auris, nasus, larynx· 0 citations
Head movements are intuitive, often spontaneous, behaviours that help us localize the origin of a sound source. In normal-hearing listeners, head movements during sound presentation are known to improve localization performance in both horizontal and vertical dimensions, as well as front-back discrimination. The percei...
T. Greif, Virginia Best, R. Baumgartner· Trends in Hearing· 0 citations
In this retrospective cohort of unilateral hearing-aid users, an REM-based fitting pathway was not associated with statistically significant 3-month audiometric or patient-reported outcome differences compared with manufacturer first-fit followed by routine fine-tuning; this does not establish equivalence or isolate th...
J. Lee, Tae Hoon Kong, J. Lee et al.· PLoS ONE· 0 citations
This study investigated the impact of visual speech cues on speech perception in noise and subjective listening effort in cochlear implant (CI) users compared with normal-hearing (NH) individuals. Furthermore, the study compared visual information presented either by a computer-generated avatar or by video recordings o...
A. Schneider, T. Stöver, T. Weissgerber· PLoS ONE· 0 citations
Although cochlear implant (CI) surgery is generally considered a safe and successful procedure, the anatomical and physiological characteristics of the inner ear may lead to certain adverse effects on the vestibular system. Subjective visual vertical (SVV) reflects the deviation in an individual’s perception of the ear...
Banu Baş, Ismail Cem Gulmez· The Egyptian Journal of Otol...· 0 citations
We use cookies to run the site and, with your consent, for analytics and to show ads.