High Dynamic Range Turbo-FLASH B1 Mapping for Characterizing Multi-Transmit Multi-Receive RF Arrays, Validated at 7 T and 11.7 T.

Sep 2026 · Magnetic Resonance in Medicine · 0 citations · 26 references
Medicine

TL;DR

This work presents a new B1 + and B1 - model-based fitting approach for interferometric presaturated TurboFLASH data, which improves B1 + and B1 - mapping robustness to noise and enables more flexible satTFL acquisition scheme, allowing better trade-offs between accuracy and scan time.

Abstract

Purpose

Transmit (B1 +) and receive (B1 -) sensitivities of parallel-transmit/receive RF array are a prerequisite for modern ultra-high-field MRI. Their accurate mapping remains challenging for coils with high transmit dynamic range, particularly in regions with low combined-mode B1 + magnitude, where standard reconstruction is impaired by noise. This work presents a new B1 + and B1 - model-based fitting approach for interferometric presaturated TurboFLASH (satTFL) data. With an optimized acquisition scheme, this method improves B1 +/B1 - mapping accuracy.

Methods

Voxel-wise B1 + and B1 - are expressed as the product of a complex unit-norm vector (phasor) and a scalar (efficiency). Phasors are estimated through singular value decomposition of satTFL signals, and efficiencies are recovered with fitting to the signal equation. This method was evaluated in phantom simulations for 7 T and 11.7 T 8Tx/32Rx RF arrays and compared with standard approaches. As it supports a variable number of pre-saturated scans, different interferometric schemes were optimized for B1 + accuracy. Finally, schemes yielding acceptable B1 + error with minimal acquisition time were tested in vitro at 7 T and 11.7 T and in vivo at 7 T.

Results

The proposed method provided lower median B1 + and B1 - errors than conventional satTFL methods and fewer outliers; the number of voxels with error > 30% is reduced from several dozens to zero. In vitro results were consistent with the simulations, and in vivo maps demonstrated good quality.

Conclusion

The proposed method improves B1 + and B1 - mapping robustness to noise and enables more flexible satTFL acquisition scheme, allowing better trade-offs between accuracy and scan time.

Read PDF

Similar papers

Open access Sep 2026

Optimizing IVIM measurement on magnetic resonance linear accelerator using multi-b-value acquisition and physics-informed neural network.

BACKGROUND Accurate quantification of intravoxel incoherent motion (IVIM) parameters on the magnetic resonance linear accelerator (MR-Linac) is challenging due to low signal-to-noise ratio and time constraints. PURPOSE This study aimed to evaluate the impact of b-value acquisition strategies and estimation algorithms...

Xin-Xiang Zhou, N. Huang, Zeng-Yi Fang et al. · 0 citations
Open access Sep 2026

Evaluations of retrospective frequency and phase correction methods for single-voxel MR spectroscopy at 7T

Subject motion and gradient heating-induced frequency and phase offsets result in spectral misalignment during single-voxel MR spectroscopy (SVS) acquisitions. Several methods have been presented to align the spectra, but their performance on a 7T system is unclear. This study aimed to evaluate the practical importance...

Chu-Yu Lee, Jia Xu, Bao-Lian Yang et al. · 0 citations
Open access Sep 2026

The Iseult Coil, a Hybrid RF Coil for the First Human Brain Images at 11.7 T: Receive Performance Comparison With a Twin Coil at 7 T

The Iseult scanner delivered the first human brain images at 11.7 T. Here, we bring forth the design of the RF coil used for these acquisitions, featuring 8 transmit and 32 receive channels, then evaluate its SNR and g‐factor advantages with respect to a similar coil at 7 T. We attempt a deduction of the SNR dependency...

Alexis Amadon, Paul-François Gapais, G. Ferrand et al. · 0 citations
Sep 2026

Acceleration-dependent variability in simultaneous T1/T2 mapping with a hybrid (mixed) sequence: a phantom comparison of SENSE, compressed sensing, and deep learning reconstructions.

BACKGROUND/PURPOSE Simultaneous T1/T2 mapping with hybrid (mixed) sequences has been proposed, but clinical use is limited by long acquisition times. Accelerated reconstruction-parallel imaging (PI; SENSE), compressed sensing (CS), and deep learning (DL)-allows shorter scans but may alter quantitative values and image...

M. Kanamoto, Yuki Matta, Toshiki Tateishi et al. · 0 citations
Open access Sep 2026

Mapping R1 of fat, R1 of water, joint R2⁎, and proton density fat fraction at 3 T using the multi-echo MP2RAGE sequence.

PURPOSE To develop a technique for joint measurement of fat and water-specific longitudinal relaxation rates (R1f and R1w), effective transverse relaxation rate (R2∗), and proton density fat fraction (PDFF) combining the Multi-Echo Magnetization Prepared Two Rapid Acquisition of Gradient Echoes (ME-MP2RAGE) sequence an...

Jorge Campos Pazmiño, Marc-Antoine Fortin, Véronique Fortier et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.