Sep 2026· Magnetic Resonance in Medicine· 0 citations· 31 references
Medicine
TL;DR
A computationally efficient image-reconstruction pipeline for 3D balanced steady state free precession CINE fetal cardiac MRI that reduces radial streaking artifacts, increases effective temporal resolution, and improves diagnostic image quality compared with previous work is developed and evaluated.
Abstract
Purpose
To develop and evaluate, in a feasibility study, a computationally efficient image-reconstruction pipeline for 3D balanced steady state free precession (bSSFP) CINE fetal cardiac MRI that reduces radial streaking artifacts, increases effective temporal resolution, and improves diagnostic image quality compared with previous work.
Methods
Three-dimensional bSSFP CINE data from 10 volunteers were used. Coil compression was performed using the region-optimized virtual coils (ROVir) technique and compared with singular value decomposition (SVD) coil compression. For ROVir, the region of interest encompassed the fetal thorax, while the region of interference was positioned over the maternal subcutaneous fat to suppress unwanted signal contributions. Image reconstruction was performed for multiple temporal samplings. Two observers evaluated diagnostic image quality using a visual-analog scale, and results were analyzed using visual grading characteristics (VGC). Computational efficiency was evaluated using reconstruction runtime and memory usage.
Results
Average 3D CINE acquisition time was 5 min, with a reconstruction time of approximately 5 h for both SVD (8channels and 8 phases) and ROVir (3 channels and 16 phases), with approximately equal memory usage. ROVir enabled higher temporal resolution by reducing the channels required. VGC analysis showed a preference for ROVir across all diagnostic criteria (AUC > 0.5), including edge sharpness, perceived image contrast, artifact suppression, papillary muscles, great vessels, and extracardiac diagnostic image quality, with comparable median scores across temporal samplings.
Conclusion
ROVir coil compression improved diagnostic quality and reduced reconstruction time. By reducing memory demands, reconstruction with a higher number of cardiac phases was also possible.
CineDiff enabled high-quality reconstruction of highly accelerated 2D RT cine CMR and reduced block-like artifacts and preserved finer anatomical detail compared with traditional compressed sensing and a variational network method, termed CineVN.
Xuan Lei, Philip Schniter, Juliet Varghese et al.· 0 citations
Motion artifacts can degrade prostate MRI quality. PROPELLER read-out (PR), a radial acquisition technique, is an established approach that can reduce motion artifacts but requires longer acquisition times. We explored whether combining PR with accelerated acquisition and reconstruction methods yields image quality...
R. Cochran, Ranjodh S. Dhami, William R. Bradley et al.· Acta Radiologica Open· 0 citations
Volumetric real-time MRI is feasible for the guidance of invasive procedures such as right-heart catheterization at 0.55T and offers flexible real-time re-slicing, volumetric 3D visualization, and the potential for improved device monitoring.
Prakash Kumar, R. Ramasawmy, A. Javed et al.· Journal of Cardiovascular Ma...· 0 citations
The RT-SVR approach enables 4D cardiac cine with comparable bi-ventricular volumes, contrast, sharpness, and acquisition time compared to standard breath-hold cine while providing additional motion robustness.
Ye Tian, A. A. Joshi, Jon A. Detterich et al.· Magnetic Resonance in Medici...· 0 citations
The SPARC pipeline (Slice-to-volume Pipeline for Automated Reconstruction of gated 3D+time fetal Cardiac MRI) which combines physics-informed slice-to-volume reconstruction (SVR) of Doppler ultrasound (DUS) gated stacks of slices, assisted by deep learning (DL) models for thoracic segmentation and anatomical reorientat...
Arnaud Boutillon, Naomi Clarke, T. Woodgate et al.· 0 citations