Skip to content
Open access

Evaluation of Implant Primary Stability Using a Standard Versus an Underpreparing Drilling Protocol with Different Measurement Methods: An In Vitro Study

Aug 2026 · Osteology · 0 citations · 20 references

TL;DR

Underpreparation appears a viable technique to improve primary stability, nevertheless it might be insufficient to reach the threshold of immediate loading suitability in low density bone.

Abstract

Background/Objectives: The primary aim of this study is to evaluate the impact of two different drilling methods on the primary stability of implants inserted into polyurethane blocks simulating different bone densities. The secondary aim is to analyse the indication provided by three different methods used to assess primary stability. Methods: The study was performed on two artificial bone blocks of solid rigid polyurethane without cortical layer and one laminated one. A total of 48 implants were placed following either a Regular Primary Stability (RPS) or an Enhanced Primary Stability (EPS) protocol based on site underpreparation. Immediately after implant placement, artificial intelligence tool (AIT) indication, peak insertion torque (pIT) and Resonance Frequency Analysis (RFA) values were recorded. The values were then categorised as suitable or not suitable for single-crown immediate loading by bone density and drilling protocol. Results: EPS drilling protocol produced significantly higher primary stability values than those in all bone densities, regardless of the primary stability measurement method used, except for the implants placed in D4 and evaluated by AIT. The grouping of values shows that implants placed in D3 and D4 never reached the immediate loading threshold for a single crown. In D2, EPS and RPS protocols both reach the threshold for all implants when evaluated with pIT or RFA; when AIT is used, RPS protocol was not able to reach the single-crown threshold in 50% of cases. Conclusions: Within the limitations of this in vitro study, underpreparation appears a viable technique to improve primary stability, nevertheless it might be insufficient to reach the threshold of immediate loading suitability in low density bone. All systems used are able to report the primary stability increase obtained by underpreparation, but AIT showed the potential to detect a loss of primary stability related to torque drop during insertion. These findings require further confirmation.

Read PDF

Similar papers

Aug 2026

The Effect of Osseodensification and Conventional Implant Site Preparation on Implant Stability of Tapered and Cylindrical Dental Implants: A Randomized Controlled Study.

Neither the choice of drilling technique nor the implant macrogeometry significantly influences insertion torque or stability levels in regions of low bone density in this randomized controlled clinical study.

Worood A Akram, S. Bede · 0 citations
Open access Jan 2026

Effect of Undersized Drilling and Sequential Osseodensification on Primary Implant Stability in Simulated Bone Types III and IV: An In Vitro Study

The use of sequential osseodensification protocols yielded superior mechanical stability parameters, specifically in IT and RT, in synthetic (polyurethane) bone models, but improvements in ISQ were predominantly noted in the denser type III model.

Silvio Yoshiaki Okagawa, A. D. de Souza, P. D. de Carvalho et al. · 0 citations
Open access Aug 2026

Influence of Different Drilling Protocols for Primary Stability at Low‐Density Bone Detected by Bone Density Measurement Drill

ABSTRACT Objectives Intraoperative bone assessment is critical for achieving primary implant stability, particularly at low‐density bone sites. Drilling torque value (DTV) has been proposed as an indicator of bone density and may help guide drilling preparation protocols to improve primary stability. However, the relat...

Yusuke Morimoto, Reiko Kobatake, K. Doi et al. · 0 citations
Open access Jul 2026

A COMPARATIVE STUDY TO ASSESS THE EFFICACY OF DIFFERENT METHODS FOR THE RETRIEVAL OF FRACTURED IMPLANT ABUTMENT SCREW: AN IN VITRO STUDY

Introduction: Fractured implant abutment screws represent a significant mechanical complication in implant dentistry and can compromise prosthetic stability, implant longevity, and treatment outcomes. Retrieval of fractured screws is a technically demanding procedure that requires effective removal while preserving imp...

Shitij Srivastava, Akshat Srivastava, A. Shekhar et al. · 0 citations
Jul 2026

Effect of Immediate Loading on the Stability of Orthodontic Mini-Implant-An Ex Vivo Study.

INTRODUCTION Mini-implant placement has become a routine method for providing anchorage in orthodontic treatment. The stability of these implants is influenced by several factors, including implant design, bone quality, site preparation, insertion angle, and loading protocol. Among these, implant and bone quality are l...

T. Chang, K. Pradeep, C. V. Padma Priya et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.