The type of scan body did not significantly affect the accuracy of digital impressions for bone-level implants and the mean and standard deviation of Root Mean Square Eror (RMSE) were calculated.
Abstract
Purpose:
Digital impression techniques in dentistry have emerged as a precise alternative to conventional methods. The accuracy of this process depends on multiple factors, including the design and material of the scan body, the type of scanner used, and the implant conditions.
Materials and Methods:
Four types of scan bodies from Dentis, Dentium, Dio, and Novodent were evaluated across 40 samples (10 per group). A reference model was first obtained with an industrial scanner (T-SCAN Hawk 2), and the samples were then scanned with an itraoral scanner (Medit i700). The resualting STL files were aligned in 3-Matic, and geometric deviations from the reference model were measured using the Best-Fit Alignment algorithm. Data were analyzed in SPSS (version 27) using one-way ANOVA with Turkey’s post hoc test, with significance set at 0.05.
Results:
The mean and standard deviation of Root Mean Square Eror (RMSE) were calculated for four scan bodies: Dentis, Dentinum, Dio, and Novodent. The reference group, Dentis, had a mean RMSE of 0.21± 0.10. Dentium showed a similar RMSE (0.21±0.08), indicating comparable accuracy to Dentis (P˃0.05). In contrast, Dio and Novodent showed lower RMSE values (0.14±0.02 and 0.15±0.06, respectively), indicating higher accuracy, although the differences were not statistically were not statistically significant (P˃0.05).
Conclusion:
The type of scan body did not significantly affect the accuracy of digital impressions for bone-level implants (P˃0.05).
Objective: This in vitro study aimed to evaluate the effect of implant vertical positioning on the accuracy of digital and conventional impressions for single implants using three-dimensional superimposition analysis.Material and Methods: A master model with a single implant analog (Biohorizons, 5.7 mm) was prepared at...
Vildan Aslıyüce, Nihal Özcan, Elif Yiğit İren et al.· Kırıkkale Üniversitesi Tıp F...· 0 citations
Evaluating how different ISB designs affect the trueness and precision of digital impressions in a completely edentulous maxillary arch with multiple implants found significant differences were found among ISBs in both trueness and precision.
Fernando Sanclemente-Reina, Carlos Serrano-Granger, Andrea Santamaría-Laorden et al.· International Dental Journal· 0 citations
Full-arch intraoral scanning of edentulous models demonstrated clinically acceptable accuracy, within the limitations of this in vitro study, and should be interpreted cautiously when extrapolating to clinical intraoral conditions.
Emel Arslan, H. Yemenoğlu· Journal of Applied Dentistry...· 0 citations
Scanner accuracy varied significantly, with Primescan and Trios 4 demonstrating superior trueness and precision and iTero and Medit showed reduced accuracy, particularly in angular measurements.
Ali Furkan Karakoyunlu, Ozum Dasdemir Ozkan· The Annals of Clinical and A...· 0 citations
The RSB technique was inferior to the desktop scan in terms of digital impression and definitive prostheses accuracy, however, the RSB technique produced zirconia prostheses within the clinically acceptable deviation value of 0.15 mm.
Ahmed Elsayyad, C. Wyatt, Ross Bryant et al.· Journal of Prosthodontics· 0 citations
Scan aids can enhance the performance of IOSs in acquiring complete-arch digital impressions, with trueness values approaching IPG systems, providing evidence for selecting optimal digital workflows in full-arch implant rehabilitation.
Lina Aljumaili, İ. Cinar· Journal of Prosthodontics· 0 citations