Diagnostic Performance of a Metasurface Coil Compared With a Commercial 12-Channel Coil for Knee MRI at 1.5 T.

Sep 2026 · Journal of Magnetic Resonance Imaging · 1 citation · 24 references
Medicine

TL;DR

The wireless metasurface coil was more convenient and provided image quality and diagnostic performance comparable to the 12-channel commercial knee coil at 1.5 T.

Abstract

Background

Knee MRI image quality depends largely on the radiofrequency coil; metasurfaces offer a new approach to improving coil performance.

Purpose

To evaluate the imaging performance and diagnostic capability of a metasurface coil for knee MRI. STUDY TYPE Prospective. PHANTOM/POPULATION A cylindrical phantom, 60 healthy volunteers (36 females; median age, 33), and 52 patients with knee injuries (28 females; median age, 47). FIELD STRENGTH/SEQUENCES T1 Weighted Spin Echo (T1W SE), Fat-Saturated Proton Density-Weighted Fast Spin-Echo (FS PDW FSE), T1-Weighted Gradient Recalled Echo (T1W GRE) at 1.5 T. ASSESSMENT All subjects underwent paired scanning with a wireless metasurface coil (plus built-in spine array) and a commercial 12-channel knee coil. Phantom metrics included signal-to-noise ratio (SNR), percent image uniformity (PIU), ghost-to-signal ratio, and B1 mapping. Volunteers were scored for image quality (5-point), SNR, contrast-to-noise ratio (CNR), and handling convenience. Patients were assessed for ACL, meniscal, and cartilage injuries against arthroscopy. STATISTICAL TESTS T-test and Wilcoxon signed-rank test for quantitative and semi-quantitative data; McNemar test for diagnostic efficacy. Kendall's coefficient of concordance and kappa, with p < 0.05 considered significant.

Results

In the phantom, SNR was 196.14 versus 183.34, PIU 92.47% versus 71.30%, and ghost-to-signal ratio 0.0093 versus 0.0145 (metasurface vs. commercial coil). Measured flip angle was 44.7° ± 1.31° (0.67% deviation from nominal 45°); B1 field uniformity was 91.14%. In volunteers, subjective image quality scores were slightly higher for the 12-channel coil on FS PDW FSE (p < 0.05), whereas scores were comparable on T1W SE and 3D FS T1W GRE (p = 0.06, 0.38). Convenience was markedly better for the metasurface (5.0 vs. 4.0, p < 0.05). In patients, diagnostic sensitivity (72.7%-83.3% vs. 72.7%-81.5%), specificity (86.7%-96.4% vs. 85.6%-96.4%), and accuracy (82.1%-90.4% vs. 81.4%-88.5%) were comparable between coils, inter-coil kappa was 0.884-1.000. "DATA"

Conclusion

The wireless metasurface coil was more convenient and provided image quality and diagnostic performance comparable to the 12-channel commercial knee coil at 1.5 T. EVIDENCE LEVEL 2. TECHNICAL EFFICACY Stage 3.

Read PDF

Similar papers

Open access Oct 2026

An Adaptive Metasurface for Adjustable Magnetic Resonance Imaging Enhancement at 3 T and 7 T: A Feasibility Study.

PURPOSE Ultra-high field (≥ 7 T) and 3 T MRI suffer from B1 + inhomogeneity and signal loss. Conventional metasurfaces that mitigate B1 + inhomogeneity and improve image SNR are "fixed" in nature and designed for a single Larmor frequency. This study demonstrates the feasibility of an adaptive metasurface utilizing vol...

Paul S. Jacobs, Pradnya Narvekar, Neil E. Wilson et al. · 0 citations
Open access Sep 2026

Computational Modelling of B₀ Field Strength Effects on Stroke MRI Image Quality: 0.5T vs 1.5T

Ischemic stroke outcomes depend on rapid, accurate neuroimaging, and magnetic resonance imaging (MRI) performance is governed by static magnetic field strength. Low-field 0.5 Tesla systems form a substantial share of installed MRI infrastructure in Nigeria and similar resource-limited settings, yet their comparative pe...

D. Amuda, O. Ogundoyin, Olanike Funke Amuda et al. · 0 citations
Open access Oct 2026

Musculoskeletal MRI Relaxometry Phantom for T1, T2 and T1rho Measurements

The phantom demonstrated excellent stability and reproducibility with temperature correction, supporting its use for quality assurance in longitudinal and multisite qMRI studies.

J. Kim, S. Russek, K. Stupic et al. · 0 citations
Open access Sep 2026

The Iseult Coil, a Hybrid RF Coil for the First Human Brain Images at 11.7 T: Receive Performance Comparison With a Twin Coil at 7 T

This study highlighted the benefits of a high B0$$ {B}_0 $$ in reception, without the potential bias of a different RF coil design; these benefits come with a slight decrease in transmission efficiency and a greater difficulty in homogenizing spin excitation.

Alexis Amadon, Paul-François Gapais, G. Ferrand et al. · 0 citations
Open access Sep 2026

Leveraging Metamaterial‐Inspired Wireless Resonator for Reliable Quantitative MRI at 1.5T: Linearity, Bias, and Repeatability Assessment

The accuracy of quantitative magnetic resonance imaging (qMRI) largely depends on the image signal‐to‐noise ratio (SNR). In recent years, studies on metamaterials in MRI have shown that they can significantly improve image quality through weak coupling with the body coil or receive coil. However, it is still unclear...

Ya-Kui Wang, Ying-Yi Qi, Bing Liu et al. · 0 citations

We use cookies to run the site and, with your consent, for analytics and to show ads. See our Cookie Policy.