Trueness and precision of 5 intraoral scanners under different conditions in arches: A 3D comparative in vitro analysis.

Oct 2026 · Dental and Medical Problems · pp. 0-0 · 0 citations
Medicine

TL;DR

While the fully dentate model showed no significant differences in trueness among the scanners, it, along with the abutment model, yielded the highest overall accuracy, whereas the edentulous and scanbody models exhibited greater deviations.

Abstract

Background

The accuracy of intraoral scanners (IOSs) has been widely reported. However, studies comparing their performance under different clinical conditions are lacking.

Objectives

The aim of the present study was to compare the trueness and precision of 5 IOSs across 4 anatomically distinct models simulating common clinical conditions: edentulous; abutment; orthodontic (fully dentate); and scanbody scenarios. MATERIAL AND

Methods

An in vitro study was conducted using 4 maxillary models representing different clinical configurations. Five IOSs (AORALSCAN 3, TRIOS 5, ViSIOScan WL, iTero Element 5D Plus, and DEXIS IS 3800) were evaluated. Each model was scanned 10 times with each scanner, and a laboratory desktop scanner was used as the reference for trueness. Accuracy was quantified using the root mean square (RMS) deviation values, with a clinical acceptability threshold of 0.25 mm. Superimposition and deviation analyses were performed using three-dimensional (3D) software, and statistical comparisons were conducted using ANOVA and Scheffé's post hoc test.

Results

Except for ViSIOScan WL in the scanbody model, all tested scanners demonstrated accuracy values below 0.1 mm. Significant differences in trueness and precision were observed both between the scanners and across the clinical scenarios (p < 0.05). Overall, AORALSCAN 3 and iTero Element 5D Plus consistently demonstrated the best performance. The abutment and fully dentate models showed the highest accuracy, whereas the edentulous and scanbody models exhibited greater deviations, particularly for ViSIOScan WL and DEXIS IS 3800.

Conclusions

Despite most IOSs achieving accuracy below 0.1 mm, substantial variability was observed depending on the scanner type and the anatomical context. AORALSCAN 3 and iTero Element 5D Plus consistently demonstrated superior trueness and precision across most models. While the fully dentate model showed no significant differences in trueness among the scanners, it, along with the abutment model, yielded the highest overall accuracy. The edentulous and scanbody models posed greater challenges.

Read PDF

Similar papers

Open access 2026

Trueness and Precision of Three Intraoral Scanners for Full-Arch Digital Impressions Exported in PLY Format: An In Vitro Study

Background: To evaluate the accuracy and precision of digital dental impressions using the Polygon File Format (PLY) across three intraoral scanners. Methods: A maxillary arch resin model with reference points was scanned 10 times using the intraoral scanners (3Shape TRIOS 5, Medit i900, Cameo Elegant 3) and a laborato...

Saud J. Dizayee · 0 citations
Open access Sep 2026

Comparative evaluation of in vivo and in vitro accuracy of intraoral scanners in splint therapy of functional disorders

The aim of this study was to conduct a comparative in vivo and in vitro assessment of the accuracy of intraoral scanners in generating a virtual dental arch model during splint therapy. The study design comprised a clinical (in vivo) phase and a laboratory (in vitro) phase. In the first phase, the trueness of scanning...

I. Naidanova, E. Bulycheva, D. S. Bulycheva · 0 citations
Open access Sep 2026

Accuracy of intraoral scanners and photogrammetry for full-arch implant-supported rehabilitations: an in vitro study

Objectives: This in vitro study aimed to compare the accuracy of intraoral scanning and photogrammetry-based acquisition methods for full-arch implant-supported rehabilitations involving four and six implants. Methods: An acrylic edentulous mandible was used to simulate the clinical setting. Implant positions were acqu...

R. Pinto, J. Martins, José Navarro et al. · 0 citations
Open access Oct 2026

Comparative Evaluation of the 3-Dimensional Implant Impression Accuracy between Conventional and Digital Impression Using Two Different Intraoral Scanners: An In Vivo Study

Background: Accurate three-dimensional (3D) recording of implant positions is fundamental to achieving passive fit and long-term success of implant-supported prostheses. The growing adoption of digital intraoral scanners (IOS) necessitates rigorous clinical evaluation of their accuracy relative to established conventio...

Rahul Ranjan, Anju Aggarwal, Punit R. S. Khurana et al. · 0 citations
Open access Sep 2026

STL Triangle Count and Trueness in Prepared Tooth Regions: Effects of Scanner System and Scanning Mode Across Intraoral, Laboratory, and Industrial Scanners.

OBJECTIVES To evaluate the effects of different scanning modes of intraoral scanners (scan-only versus restoration type selected) on STL triangle counts and to assess the influence of STL triangle counts on trueness across different scanner systems. METHODS A typodont model with 3 prepared teeth was scanned with 15 s...

Hale Arikan Kalayci, Simge Alıcı, Gülçe Çakmak et al. · 0 citations
Sep 2026

Influence of implant scan body material and intraoral scanner system on the accuracy of three-unit implant scans.

Within the limitations of this in vitro study, iTL demonstrated higher trueness and precision than the other IOSs tested, whereas PEEK and hybrid ISBs demonstrated higher trueness for 3-unit implant scans.

Bahar Gurpinar, Burcu Kanat-Ertürk, Mehmet Karakuzu et al. · 0 citations